AC Voltage and AC Current, DC Volts, DC Current,
View all Fluke 8845A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 30 highlights
8845A/8846A Users Manual Diode Test Test Current 100 µA or 1 mA Response Time 300 samples/sec with audible tone. Accuracy is given as ± (% measurements + % of range) Range 5.0000 V 10.0000 V 24 Hour (23 ± 1 °C) 0.002 + 0.002 0.002 + 0.001 90 Days (23 ± 5 °C) 0.008 + 0.002 0.008 + 0.002 1 Year (23 ± 5 °C) 0.01 + 0.002 0.01 + 0.002 Temperature Coefficient/ °C Outside 18 to 28 °C 0.001 + 0.002 0.001 + 0.002 Measurement Rates Function Digits Integration Time Setting 60 Hz (50 Hz) Measurements/Second[1] 8845A 8846A DC Volts, DC Current, and Resistance 6½ 100 NPLC 1.67 (2) s 0.6 (0.5) 6½ 10 NPLC 167 (200) ms 6 (5) 0.6 (0.5) 6 (5) 5½ 10 NPLC 16.7 (20) ms 60 (50) 60 (50) 5½ 0.2 NPLC 3 ms 300 300 AC Voltage and AC Current[2] 6½ 3 Hz 0.14 0.14 6½ 20 Hz 6½ 200 Hz[3] 6½ 200 Hz[4] 1 1 1.6 1.6 6 6 Frequency and Period 6½ 1 s 1 1 5½ 100 ms 9.8 9.8 4½ 10 ms 80 80 [1] Typical measurement rates with autozero off. [2] Maximum measurement rates for 0.01 % of ac step. When dc input varies, additional settling delay is required. [3] For remote operation or external trigger using default settling delay [4] Settling delay = 0 1-20