Fluke 8845A FE 8845A & 8846A Users Manual - Page 35

Current-Input Fuses, Table 2-1. Line Voltage to Fuse Rating, Line Voltage Selection, Fuse Rating

Page 35 highlights

2 Preparing the Meter for Operation Power Considerations Table 2-1. Line Voltage to Fuse Rating Line Voltage Selection Fuse Rating Fluke Part No. 100 0.25 A, 250 V (slow blow) 120 0.25 A, 250 V (slow blow) 220 0.125 A, 250 V (slow blow) 240 0.125 A, 250 V (slow blow) 166306 166306 166488 166488 120 Figure 2-1. Replacing the Line Fuse caw0201f.eps Current-Input Fuses The 100 mA and 10 A inputs are protected by user-replaceable fuses. • The 100 mA input is protected by a fuse (F2) rated at 440 mA, 1000 V (fast blow), 10,000 A minimum breaking capacity (Fluke Part No. 943121). • The 10 A input is protected by a fuse (F1) rated at 11 A, 1000 V (fast blow), 10,000 A minimum breaking capacity (Fluke Part No. 803293). W Warning For protection against fire or arc flash, replace a blown fuse only with one from Fluke. To test for a blown Current Input Fuse: 1. With the Meter powered up, plug a test lead into the VΩGER connector. 2. Press N. 3. Insert the other end of the test lead into the 100 mA input connector. If the fuse is good, the Meter will read less than 200 Ω. If the fuse is blown, the Meter will read overload. 4. Remove the probe from the 100 mA connector and insert into the 10 A connector. 2-5

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93

Preparing the Meter for Operation
Power Considerations
2
2-5
Table 2-1. Line Voltage to Fuse Rating
Line Voltage Selection
Fuse Rating
Fluke Part No.
100
0.25 A, 250 V (slow blow)
166306
120
0.25 A, 250 V (slow blow)
166306
220
0.125 A, 250 V (slow blow)
166488
240
0.125 A, 250 V (slow blow)
166488
120
caw0201f.eps
Figure 2-1. Replacing the Line Fuse
Current-Input Fuses
The 100 mA and 10 A inputs are protected by user-replaceable fuses.
The 100 mA input is protected by a fuse (F2) rated at 440 mA, 1000 V (fast
blow), 10,000 A minimum breaking capacity (Fluke Part No. 943121).
The 10 A input is protected by a fuse (F1) rated at 11 A, 1000 V (fast blow),
10,000 A minimum breaking capacity (Fluke Part No. 803293).
W
Warning
For protection against fire or arc flash, replace a blown fuse
only with one from Fluke.
To test for a blown Current Input Fuse:
1.
With the Meter powered up, plug a test lead into the
V
GER
connector.
2.
Press
N
.
3.
Insert the other end of the test lead into the 100 mA input connector.
If the fuse is good, the Meter will read less than 200
. If the fuse is blown, the
Meter will read
overload
.
4.
Remove the probe from the 100 mA connector and insert into the 10 A
connector.