Fluke 8845A FE 8845A & 8846A Users Manual - Page 35
Current-Input Fuses, Table 2-1. Line Voltage to Fuse Rating, Line Voltage Selection, Fuse Rating
View all Fluke 8845A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 35 highlights
2 Preparing the Meter for Operation Power Considerations Table 2-1. Line Voltage to Fuse Rating Line Voltage Selection Fuse Rating Fluke Part No. 100 0.25 A, 250 V (slow blow) 120 0.25 A, 250 V (slow blow) 220 0.125 A, 250 V (slow blow) 240 0.125 A, 250 V (slow blow) 166306 166306 166488 166488 120 Figure 2-1. Replacing the Line Fuse caw0201f.eps Current-Input Fuses The 100 mA and 10 A inputs are protected by user-replaceable fuses. • The 100 mA input is protected by a fuse (F2) rated at 440 mA, 1000 V (fast blow), 10,000 A minimum breaking capacity (Fluke Part No. 943121). • The 10 A input is protected by a fuse (F1) rated at 11 A, 1000 V (fast blow), 10,000 A minimum breaking capacity (Fluke Part No. 803293). W Warning For protection against fire or arc flash, replace a blown fuse only with one from Fluke. To test for a blown Current Input Fuse: 1. With the Meter powered up, plug a test lead into the VΩGER connector. 2. Press N. 3. Insert the other end of the test lead into the 100 mA input connector. If the fuse is good, the Meter will read less than 200 Ω. If the fuse is blown, the Meter will read overload. 4. Remove the probe from the 100 mA connector and insert into the 10 A connector. 2-5