Konica Minolta bizhub PRESS C71hc bizhub PRESS C1060/C1070/C1070P/PRO C1060L D - Page 62
Select [Spot] on [Measurement Mode]
View all Konica Minolta bizhub PRESS C71hc manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 62 highlights
14 Select [Spot] on [Measurement Mode], and click [Calibrate]. 15 When the following message is displayed, horizontally place the spectrophotometer on the calibration dock and click [Calibrate]. 16 Click [Close]. 17 Stack 10 sheets of blank paper for the maximum density management, and place the job that is output in step 2 on top of it. 18 Measure all the maximum density patches of K. Place the spectrophotometer on the patch, and press the button on the spectrophotometer. When you hear a beep sound, check that the read patch is displayed on the screen. Tips For i1Pro2/ES-2000, install the positioning target to the spectrophotometer before you measure the color. 3-30
3-30
14
Select [Spot] on [Measurement Mode],
and click [Calibrate].
15
When the following message is displayed, horizontally place the spectrophotometer
on the calibration dock and click [Calibrate].
16
Click [Close].
17
Stack 10 sheets of blank paper for the maximum density management, and place the
job that is output in step 2 on top of it.
18
Measure all the maximum density
patches of K.
Place the spectrophotometer on the patch, and
press the button on the spectrophotometer. When
you hear a beep sound, check that the read patch
is displayed on the screen.
Tips
For
i1Pro2/ES-2000
, install the positioning target to the
spectrophotometer before you measure the color.