Konica Minolta bizhub PRESS C71hc FD-9 Auto Scan Spectrophotometer User Guide - Page 33
condition them., Wipe off the rubber rollers with alcohol to clean
View all Konica Minolta bizhub PRESS C71hc manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 33 highlights
1 Make sure that the FD-9 is switched off. 2 Disconnect all cables (USB cable, LAN cable, AC adapter cable). 3 Open the cover. 4 Loosen the two screws as shown completely so that they are free of the base. 5 Close the cover. 6 Grasp the ends of the FD-9 and tilt the FD-9 body backward as shown. The base should stay flat. 7 Clean the cover glass of the line sensor and the white backing and other plastic surfaces with a clean, dry, soft, lint-free cloth. 8 Wipe off the rubber rollers with alcohol to clean and condition them. 9 Grasp the ends of the FD-9 and tilt the FD-9 body onto the base again. 10 Open the cover 11 Tighten the two screws as shown until snug. • If the screws are loose when measurements are taken, measurement accuracy may suffer. • Do not overtighten. 12 Close the cover and reconnect cables. Cleaning 31