Sony DVO1000MD User Manual (Instructions For Use) - Page 5

Guidance and manufacturer's declaration - electromagnetic immunity, > 95% dip

Page 5 highlights

Guidance and manufacturer's declaration - electromagnetic immunity The DVO-1000MD is intended for use in the electromagnetic environment specified below. The customer or the user of the DVO-1000MD should assure that it is used in such as environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact ±8 kV air ±6 kV contact ±8 kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. Electrical fast transient/burst ±2 kV for power supply lines ±2 kV for power supply lines Mains power quality should be that of a typical commercial or hospital environment. IEC 61000-4-4 ±1 kV for input/ output lines Surge IEC 61000-4-5 ±1 kV differential mode ±2 kV common mode Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 < 5% UT (> 95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles ±1 kV for input/ output lines ±1 kV differential mode ±2 kV common mode < 5% UT (> 95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the DVO-1000MD requires continued operation during power mains interruptions, it is recommended that the DVO-1000MD be powered from an uninterruptible power supply or a battery. 70% UT (30% dip in UT) for 25 cycles 70% UT (30% dip in UT) for 25 cycles Power frequency (50/60Hz) magnetic field < 5% UT (> 95% dip in UT) for 5 sec 3 A/m < 5% UT (> 95% dip in UT) for 5 sec 3 A/m Power frequency magnetic fields should be at least characteristic of a typical location in a typical commercial or hospital environment. IEC 61000-4-8 NOTE: UT is the a.c. mains voltage prior to application of the test level. 5

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69

5
Guidance and manufacturer’s declaration - electromagnetic immunity
The DVO-1000MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
DVO-1000MD should assure that it is used in such as environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5%
U
T
(> 95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
)
for 5 sec
< 5%
U
T
(> 95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
< 5%
U
T
(> 95% dip in
U
T
)
for 5 sec
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the DVO-1000MD
requires continued operation during power mains
interruptions, it is recommended that the DVO-1000MD be
powered from an uninterruptible power supply or a battery.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at least
characteristic of a typical location in a typical commercial or
hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.