Fluke Ti55FT-20 User Manual - Page 109
Table D-2. TiR3, TiR4, Ti55 and Ti50 Specifications cont., Model Name, Ti55 and TiR4, Ti50 and TiR3
View all Fluke Ti55FT-20 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 109 highlights
D Appendices Camera Specifications and Dimensions Table D-2. TiR3, TiR4, Ti55 and Ti50 Specifications (cont.) Model Name Ti55 and TiR4 Ti50 and TiR3 Calibrated Temperature Ranges Accuracy Measurement Modes Temperature Measurement Range 1 = -20 °C to 100 °C Range 1 = -20 °C to 100 °C (-4 °F to 212 °F) (-4 °F to 212 °F) Range 2 and 3 are not available on the TiR4 Range 2 is not available on the TiR3 Range 2 = -20 °C to 350 °C Range 2 = -20 °C to 350 °C (-4˚F to 662˚F) (-4 °F to 662 °F) Range 3 = 250 °C to 600 °C (482 °F to 1112 °F) ±2 °C or ±2 %, whichever is greater Movable points & boxes Centerpoint Cursor temp. Center box (area min/max, average) Automatic hot and cold point detection Color alarm Centerpoint Cursor temp. Center box (area min/max, average) Emissivity Correction Battery Type Based on user input. Variable from 0.01 to 1.0 (0.01 increments) Power Li-Ion Smart Battery, rechargeable, field-replaceable Battery Oper. 2-3 hours continuous operation for each battery, typical Time D-9