Fluke 753 Fluke 750 Series User Manual - Page 43
Continuity Test, Pressure Measurement, To prevent personal injury, shut off the valve
View all Fluke 753 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 43 highlights
Continuity Test When you do a continuity test, the beeper sounds and Short is shown on the display when the resistance between the Ω MEASURE jack and its common jack is less than 25 Ω. Open is shown when the resistance is larger than 400 Ω. To do a continuity test: 1. De-energize the circuit under test. 2. If necessary, push for MEASURE mode. 3. Push q twice so that Open is shown. 4. Connect the Product to the circuit under test. See Figure 12. Documenting Process Calibrator Measure Mode Pressure Measurement Many ranges and types of pressure modules are available from Fluke. See "Accessories". Before you use a pressure module, read its instruction sheet. The modules are different in how they are used, zeroed, what types of process pressure media are allowed, and accuracy specifications. Figure 13 shows gage and differential modules. Differential modules also operate in gage mode when you leave the low fitting open to atmosphere. To measure pressure, attach the applicable pressure module for the process pressure you will test as described in the module's Instruction Sheet. To measure pressure: Warning To prevent personal injury, shut off the valve and slowly bleed off the pressure before attaching the pressure module to the pressure line to avoid a violent release of pressure in a pressurized system. 31