Next Pressure Units, Pressurize the pressure line with the pressure source
View all Fluke 753 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 64 highlights
753/754 Users Manual 1. Connect a pressure module and pressure source to the Product as Figure 21 shows. The threads on the pressure modules accept ¼ NPT fittings. Use the supplied ¼ NPT to ¼ ISO adapter if necessary. 2. If necessary, push for SOURCE mode. 3. Push p. The Product automatically senses which pressure module is attached and sets its range accordingly. 4. Zero the pressure module as described in the module's instruction sheet. The module types are different in how they are zeroed. The pressure module MUST be zeroed before doing a task that sources or measures pressure. 5. Pressurize the pressure line with the pressure source to the necessary level as shown on the display. 6. If necessary, change the pressure display units to psi, mHg, inHg, inH2O, ftH2O, mH2O, bar, Pa , g/cm2, or inH2O@60°F. Metric units (kPa, mmHg, etc.) are shown in Setup mode in their base units (Pa, mHg, etc.). To change the pressure display units: 1. Push s. 2. Push Next Page twice. 3. Push with the cursor on Pressure Units. 4. Select the pressure units with or . 5. Push . 6. Push the Done softkey. 52