HP Indigo WS6600 Dripping Prevention and ITM Heating Hood Service -- CA393-101 - Page 63

Press in Ready status, Main Menu > PIP and Blanket > First Transfer

Page 63 highlights

1. In Systab Editor, set IO_COOLER_FAN to 10000. 2. Place the Anemometer at the outlet duct as shown below. The vane should be in the middle of the utility cabinet outlet duct. Air velocity should be 5 ± 2 m/sec. Figure 3-43 Utility cabinet outlet Press in Ready status Floow these stteps for dripping correction procedures when the press in ready status. 1. Put the press into Ready status. 2. Check the machine temperature control as follows. With the press in Ready (print) status, touch Main Menu > Press status > System status, and verify that temperature values are within specifications. Figure 3-44 System status showing machine temperature 3. In Systab Editor, IO_Outlet_Temperature_Sensor, the value should be between 20°C and 24°C. 4. Using a lint-free wipe with imaging oil, wipe the entire PIP surface, including the PIP shoulders. 5. Run the First transfer calibration: Main Menu > PIP and Blanket > First Transfer. 6. Perform the following test procedure: 60 Chapter 3 Troubleshooting

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72

1.
In Systab Editor, set
IO_COOLER_FAN
to
10000
.
2.
Place the Anemometer at the outlet duct as shown below. The vane should be in the middle of the
utility cabinet outlet duct.
Air velocity should be 5 ± 2 m/sec.
Figure 3-43
Utility cabinet outlet
Press in Ready status
Floow these stteps for dripping correction procedures when the press in ready status.
1.
Put the press into
Ready
status.
2.
Check the machine temperature control as follows.
With the press in Ready (print) status, touch
Main Menu > Press status > System status
, and verify
that temperature values are within specifications.
Figure 3-44
System status showing machine temperature
3.
In Systab Editor,
IO_Outlet_Temperature_Sensor
, the value should be between 20°C and 24°C.
4.
Using a lint-free wipe with imaging oil, wipe the entire PIP surface, including the PIP shoulders.
5.
Run the First transfer calibration:
Main Menu > PIP and Blanket > First Transfer
.
6.
Perform the following test procedure:
60
Chapter 3
Troubleshooting