Invacare IFX-20MP Owners Manual - Page 19
Iec 60601 Test, Immunity Test
View all Invacare IFX-20MP manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 19 highlights
3 Electromagnetic Compatibility (EMC) Information Power frequency (50/60 Hz) 3 A/m magnetic field IEC 61000-48 30 A/m NOTE: UT is the a.c. mains voltage prior to application of the test level. Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Guidance and manufacturer's declaration - electromagnetic immunity The AVIVA FX Power Wheelchair is intended for use in the electromagnetic environment specified below. The customer or the user of the AVIVA FX Power Wheelchair should assure that it is used in such an environment. IMMUNITY test IEC 60601 TEST LEVEL Compliance level Electromagnetic environment - guidance Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3 3 Vrms 150 kHz to 80 MHz 3 V/m 80 MHz to 2,5 GHz 3 Vrms 20 V/m 26 MHz to 2.5 GHz per ISO 7176-21 Portable and mobile RF communications equipment should be used no closer to any part of the AVIVA FX Power Wheelchair, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance (3.5/ 3 V/m = 1.667) 26 MHz to 800 MHz (3.5/ 20 V/m = 0.175) 60101916-A 800 MHz to 2,5 GHz (7/ 20V/m = 0.35) 19