Invacare IFX-20MP Owners Manual - Page 61
How The User is Secured Within the Mobility Device
View all Invacare IFX-20MP manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 61 highlights
Fig. 8-2 Seat Tie-Down Points-Wheelchairs with Ultra Low Maxx Seats and Elevate 1. Secure the mobility device with the tie-down system belts at the following locations: All Wheelchairs - four tie down points C on the wheelchair base (two at the front and two at the rear). Systems with Ultra Low Maxx Seats and Elevate (TRRO-E) - two additional tie down points B on either side of the seat at the front of the seat rails. 8 Transport - ISO 7176-19 2. Secure the mobility device by tensioning the belts in accordance with the tie-down system manufacturer's user manual. 8.5.2 How The User is Secured Within the Mobility Device CAUTION! Risk of injury if the user is not properly secured within the mobility device -Even if the mobility device is equipped with a posture belt, this is no substitute for a proper safety belt which complies with ISO 10542 in the transport vehicle. Always use the safety belt installed in the transport vehicle. -Safety belts must be in contact with the user's body. They must not be held at a distance from the user's body using parts of the mobility device such as armrests or wheels. -Safety belts must be pulled as tightly as possible without causing the user discomfort. -Safety belts must not be positioned while twisted. -Ensure that the third seat belt anchorage point is not fixed directly to the vehicle floor, but to one of the vehicle uprights. 60101916-A 61