Invacare IFX-20MP Owners Manual - Page 80
Removing the Batteries from the Wheelchair, Using the Battery Isolation Switch,
View all Invacare IFX-20MP manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 80 highlights
Invacare® AVIVA™ FX 10.5.4 Removing the Batteries from the Wheelchair CAUTION! Risk of Damage Cleaning or maintenance may cause damage to carpeting or flooring. -Place the wheelchair in a well ventilated area where cleaning or maintenance can be performed without risk of damage to carpeting or flooring. l 10 mm Hex Socket 1. Place the wheelchair in a well ventilated area. 2. Verify the battery isolation switch is in the Off position. Refer to 7.3 Using the Battery Isolation Switch, page 48 for more information. 3. Remove the rear shroud assembly. Refer to 10.4.1 Removing/Installing the Rear Shroud Assembly , page 75. 4. Disconnect the left motor cable G from the power module P. Fig. 10-5 Disconnecting the motor cable 5. Using the 10 mm socket, remove the two bolts M securing the battery door assembly N to the rear of the battery box O. 80 60101916-A