Invacare M41SR16R Owners Manual - Page 28
Troubleshooting - Electrical
View all Invacare M41SR16R manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 28 highlights
SECTION 5-SAFETY INSPECTION Troubleshooting - Electrical NOTE: For additional troubleshooting information and explanation of error codes, refer to the individual Electronics Manual supplied with each wheelchair. SPJ+ Joysticks The joystick information gauge and the service indicator give indications of the type of fault or error detected by the control module. When a fault is detected, the wheelchair may stop and not drive. The LEDs on the information gauge may flash in a particular pattern or the service indicator light will flash. The number or type of flashes indicates the nature of the error. If multiple errors are found, only the first error encountered by the control module will be displayed. Information Gauge Display Diagnostics DISPLAY Information Gauge Display DESCRIPTION DEFINITION COMMENTS All LEDs are off. Power is off. All LEDs are on. Power is on. Fewer than three LEDs on implies reduced battery charge. Left RED LED is flashing. Battery charge is low. The batteries should be charged as soon as possible. Left to Right "chase" alternating with steady display. Joystick is in The steady LEDs indicate programming, inhibit and/ the current state of the or charging mode. battery charge. All LEDs are flashing slowly. Joystick has detected Out- Release the joystick back of-Neutral-at-Power-Up to Neutral. mode. Pronto® M41 with SureStep® 28 Part No. 1143206