Panasonic BT-4LH310 Parts List - Page 54

Procedure of writing LCD Panel data, Gamma Corrector_BT4LH310.EXE

Page 54 highlights

1-2. Procedure of writing LCD Panel data Please write LCD Panel data as follows. 1. Boot up the unit in the service mode (While simultaneously pressing the"SDI","HDMI-DUAL", "HDMI-SINGLE" button, turn ON the power.). 2. Press the MENU button to open the service menu and confirm that the item "REMOTE TYPE" set to "ACK/CS OFF". 3. Turn the power off and then on again. (Boot up the unit in the normal mode). 4. Double-click the file "GammaCorrector_BT-4LH310.exe" to boot up the gamma adjustment software. Gamma Corrector_BT4LH310.EXE NOTE: When the ERROR message appears, take the following procedures. 1. Click the "OK" button in the ERROR message so that the Gamma Corrector screen appears. 2. Click the "File" → "Property(P)" button so that the Property screen appears. 3. Set the Analyzer MODEL to "No Connect" and click the "OK" button. 4. Restart the Gamma corrector application. 5. BT-4LH310 Gamma Corrector screen appears as follows. ADJ - 3

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125

ADJ - 3
1-2. Procedure of writing LCD Panel data
Please write LCD Panel data as follows.
1. Boot up the unit in the service mode (While simultaneously pressing the
"SDI","HDMI–DUAL", "HDMI–SINGLE"
button,
turn ON the power.).
2. Press the
MENU
button to open the service menu and confirm that the item "
REMOTE TYPE"
set to "
ACK/CS OFF"
.
3. Turn the power off and then on again. (Boot up the unit in the normal mode).
4. Double-click the file "GammaCorrector_BT-4LH310.exe" to boot up the gamma adjustment software.
NOTE: When the ERROR message appears, take the following procedures.
1.
Click the
"OK"
button
in the ERROR message so that the Gamma Corrector screen appears.
2.
Click the
"File"
"Property(P)"
button so that the Property screen appears.
3.
Set the Analyzer MODEL to "
No Connect"
and click the "
OK"
button.
4.
Restart the Gamma corrector application.
5.
BT-4LH310 Gamma Corrector screen
appears as follows.
Gamma Corrector_BT4LH310.EXE