Panasonic BT-4LH310 Parts List - Page 57
Name of each part CA-210, 4. Calibration for BT-4LH310, MEMORY CH key
View all Panasonic BT-4LH310 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 57 highlights
2-3. Name of each part (CA-210) Measurement mode display part White key LCD display part MENU key CAL/SPACE key ENTER key 0-CAL key MODE key HOLD key Red key MEMORY CH key Green key Bluekey 2-4. Calibration for BT-4LH310 Calibration for BT-4LH310 is required for the first connecting with CA-210 and BT-4LH310 to execute the gamma measurement. Once the calibration values are stored in CA-210, the calibration for BT-4LH310 need not be executed every ime. 2-4-1 Preparation for calibration Service Web Site and unzip it. Please perform the aging of BT-4LH310 for one hour or more. 1. When calibration is executed, aging (for one hour or more after turning the power on) is required. NOTE: While simultaneously pressing the "FUNCTION 4" button and the "FUNCTION 5" button, turn ON the power to indicate the aging time on the left upper corner of the screen. (Please hold down the buttons until the character appears on the screen.) 14M Minutes 1. Press the "MENU" button to open the Main Menu. 2. Open the SYSTEM CONFIG menu screen and set the item "POWER SAVE MODE" to "OFF". ADJ - 6