Fluke TI450 SF6 User Manual - Page 34
Background (Reflected Background Temperature Compensation), Select by Table
View all Fluke TI450 SF6 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 34 highlights
Ti200, Ti300, Ti400,Ti450 Users Manual Select by Table To select from a list of common materials: 1. Go to Measurement > Emissivity > Select Table. 2. Push / to highlight the material. 3. Push to select the material. Background (Reflected Background Temperature Compensation) Compensation for reflected background temperature is set in the Background tab. Very hot objects or very cold objects can affect the apparent temperature and measurement accuracy of the target or object of interest, especially when surface emissivity is low. Adjustment of the reflected background temperature can make the temperature measurement better in many situations. For more information, see Emissivity Adjustment. To adjust the background temperature: 1. Go to Measurement > Background. 2. Push / to change the value. 3. Push or when done. Note If the Display is set to Display All, you see the information about current reflected background temperature as BG = xx.x. 30