Fluke TI450 SF6 User Manual - Page 35
Transmission/Transmittance Adjustment, Spot Temperatures, Thermal Imagers, Display All
View all Fluke TI450 SF6 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 35 highlights
Thermal Imagers Menus Transmission/Transmittance Adjustment When you do infrared inspections through infrared-transparent windows (IR windows), not all of the infrared energy emitted from the objects of interest is transmitted through the optical material in the window. If the transmission percentage of the window is known, you can adjust this percentage in the Imager or in the SmartView software. Adjustment of the transmission correction can make the accuracy of the temperature measurement better in many situations. To adjust the transmission percentage: 1. Go to Measurement > Transmission. 2. Push / to adjust the percentage between 10 % and 100 %. 3. Push or when done. Note If Display Information is set to Display All, you see the information about current transmission correction as τ = xxx%. Spot Temperatures The Spot Temperatures are floating HI and LO temperature indicators that move on the display as the temperature measurements of the image fluctuate. To turn on/off the hot and cold spot indicators: 1. Go to Measurement > Spot Temp. 2. Push / to highlight On or Off. 3. Push or to set the new value. 31