Measuring Silver ink OD

Page 36 highlights

Measuring Silver ink OD ● Silver ink OD should be measured in M3 conditions (with polarizers). This can be done by the external eXact spectrophotometer. ● The on-press spectrophotometer (SPM) can measure only in M0 conditions. ● In order to measure the OD of the silver ink with an M0 external spectrophotometer, you need to measure the OD of Silver printed by 2 hits, subtracted by the OD of Silver printed over 1 hit of black, subtracted by the OD of 2 hits of silver = 0.07 ± 0.02 in black channel (for coated gloss and semi matte - white substrates only). ● The M0 output is the subtraction of 2 elements: Target OD of Silver (M0) = OD of 2 hits of Silver on Black - OD of 2 hits of Silver NOTE: ● M3 is higher ⇒ M0 is lower ⇒ more Silver ● M3 is lower ⇒ M0 is higher ⇒ less Silver ● Typical values for M3: 0.9 -1.3 ● OD target range for M0: 0.05 - 0.16 32 Appendix A Full color calibration

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45

Measuring Silver ink OD
Silver ink OD should be measured in M3
conditions (with polarizers). This can be done
by the external eXact spectrophotometer.
The on-press spectrophotometer (SPM) can
measure only in M0 conditions.
In order to measure the OD of the silver ink with
an M0 external spectrophotometer, you need
to measure the OD of Silver printed by 2 hits,
subtracted by the OD of Silver printed over 1 hit
of black, subtracted by the OD of 2 hits of silver
= 0.07 ± 0.02 in black channel (for coated gloss
and semi matte – white substrates only).
The M0 output is the subtraction of 2 elements:
Target OD of Silver (M0) = OD of 2 hits of Silver on Black – OD of 2 hits of Silver
NOTE:
M3 is higher
M0 is lower
more Silver
M3 is lower
M0 is higher
less Silver
Typical values for M3: 0.9 -1.3
OD target range for M0: 0.05 – 0.16
32
Appendix A
Full color calibration