Campbell Scientific IRGASON IRGASON Integrated CO2/H2O Open-Path Gas Analyzer - Page 14
Sonic temperature, Full scale wind
![]() |
View all Campbell Scientific IRGASON manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
IRGASON® Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer Sonic accuracy5 Offset error ux, uy: uz: < ±8 cm·s-1 < 4 cm·s-1 Gain error Wind vector ±5° horizontal: < ±2% of reading Wind vector ±10° horizontal: < ±3% of reading Wind vector ±20° horizontal: < ±6% of reading Sonic reporting range Full scale wind: Sonic temperature: ±65.553 m/s -50° to +60°C Auxiliary sensors6 Barometer Basic barometer Accuracy -30 to 0°C: ±3.7 kPa at -30°C, falling linearly to ±1.5 kPa at 0oC 0° to 50°C: ±1.5 kPa Measurement rate: 10 Hz Optional enhanced barometer: Manufacturer: Vaisala Model: PTB110 Accuracy -30 to 0°C: ±0.15 kPa Measurement rate: 1 Hz IRGASON temperature sensor: Manufacturer: BetaTherm Model: Accuracy: 100K6A1A Thermistor ±0.15oC (-30o to 50oC) 1 noise rms, assumes: o 25°C o 85 kPa o 14 g·m-3 H2O o 597 mg·m-3 CO2 o 25 Hz bandwidth. 2 user selectable 3 -30° to 50°C 4 noise rms 5 assumes: o -30° to +50°C o wind speed
![](/manual_guide/products/campbell-scientific-irgason-irgason-integrated-co2h2o-openpath-gas-analyzer-3d-sonic-anemometer-0d79b50/14.png)