Campbell Scientific IRGASON IRGASON Integrated CO2/H2O Open-Path Gas Analyzer - Page 46
TABLE 10-3., Bits in the Gas Diagnostic Flag, hex value, decimal, Function
View all Campbell Scientific IRGASON manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 46 highlights
IRGASON® Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer TABLE 10-3. Bits in the Gas Diagnostic Flag bit hex value decimal 0 0x1 1 1 0x2 2 2 0x4 4 3 0x8 8 4 0x10 16 5 0x20 32 6 0x40 64 7 0x80 128 8 0x100 256 9 0x200 512 10 0x400 1024 11 0x800 12 0x1000 13 0x2000 14 0x4000 15 0x8000 16 0x10000 2048 4096 8192 16384 32768 65536 17 0x20000 131072 18 0x40000 262144 19 0x80000 524288 20 0x100000 1048576 21 0x200000 2097152 22 0x400000 4194304 Name Function Bad Data Data are suspect (there is an active diagnostic flag) Sys Fault General system fault Sys Startup Gas analyzer is starting up Motor Speed Motor speed outside of limits TEC Temp TEC temperature exceeds limits Light Power Source power exceeds limits Light Temp Invalid source temperature Light I Source current exceeds limits Power Off Gas head not powered Chan Err Gas input data out of sync with home pulse Amb Temp Invalid ambient temperature Amb Press Invalid ambient pressure CO2 I CO2 I exceeds limits CO2 Io CO2 Io exceeds limits H2O I H2O I exceeds limits H2O Io H2O Io exceeds limits CO2 Io Var Moving variation in CO2 Io exceeds limits H2O Io Var Moving variation in H2O Io exceeds limits CO2 Io Ratio CO2 signal level too low H2O Io Ratio H2O signal level too low Cal Mem Err Gas head calibration signature error Heater Control Heater control error Diff Pressure Differential pressure exceeds limits 36