Campbell Scientific IRGASON IRGASON Integrated CO2/H2O Open-Path Gas Analyzer - Page 57
TABLE A-1., Filter Time Delays for Various Bandwidths, Bandwidth Hz, Time Delay ms, Appendix A.
View all Campbell Scientific IRGASON manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 57 highlights
No Units Appendix A. Filter Bandwidth and Time Delay EC100 10-Hz Filter Compared to 20-msec Moving Average (Amplitude Responses) 10 1 0.1 0.01 EC100 10-Hz Bandwidth Filter 10-Hz Bandwidth from a 50-msec Moving Average 0.001 0.0001 1 10 70 Hertz FIGURE A-2. Frequency response comparison of the EC100 10 Hz bandwidth and a 50 ms moving average TABLE A-1. Filter Time Delays for Various Bandwidths Bandwidth (Hz) 5 10 12.5 20 25 Time Delay (ms) 800 400 320 200 160 A-3
Appendix A.
Filter Bandwidth and Time Delay
FIGURE A-2.
Frequency response comparison of the EC100 10 Hz
bandwidth and a 50 ms moving average
TABLE A-1.
Filter Time Delays for Various Bandwidths
Bandwidth (Hz)
Time Delay (ms)
5
800
10
400
12.5
320
20
200
25
160
1
10
70
Hertz
0.0001
0.001
0.01
0.1
1
10
No Units
EC100 10-Hz Bandwidth Filter
10-Hz Bandwidth from a 50-msec Moving Average
EC100 10-Hz Filter Compared to 20-msec Moving Average (Amplitude Responses)
A-3