Campbell Scientific IRGASON IRGASON Integrated CO2/H2O Open-Path Gas Analyzer - Page 27
Unprompted Output, 7.6 RS-485 Baud Rate, 7.7 Analog Output, 7.8 ECMon Update Rate
View all Campbell Scientific IRGASON manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 27 highlights
IRGASON® Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer any integer value between 0 and 14. The value 15 is reserved as an SDM-group trigger. 7.4 Unprompted Output If the EC100 is to output data as USB or RS-485 (see Section 8.2, USB or RS-485 Output), this setting must be set to USB Port or RS485, respectively. The factory default is to disable the unprompted output and assume data are to be logged via SDM (see Section 8.1, SDM Output). Only one unprompted-output type (i.e., USB or RS-485) may be selected at a given time. The rate at which the EC100 outputs these data is determined by the Unprompted Output Rate setting. 7.5 Unprompted Output Rate This setting determines the output rate for unprompted output (USB or RS-485; see Section 8.2, USB or RS-485 Output). If the unprompted output is disabled, this parameter is not used. The factory-default output rate is 10 Hz, but it may be set to 10, 25, or 50 Hz. 7.6 RS-485 Baud Rate If the unprompted output mode is set to RS-485, this parameter determines the baud rate. Otherwise, this setting is not used. The RS-485 baud rate defaults to 115200 bps, although the user may enter another value. 7.7 Analog Output The EC100 has two analog outputs for CO2 and H2O densities (see Section 8.3, Analog Outputs, for more information). These outputs may be enabled or disabled with this setting. The default is for analog output to be disabled. Note that the analog output does not include sonic data. 7.8 ECMon Update Rate This setting determines the rate at which data are sent over the USB connection to the PC while running ECMon. The default setting of 10 Hz should be adequate in most situations. 7.9 Temperature Sensor This setting configures the EC100 electronics to integrate with the IRGASON. The IRGASON measures ambient air temperature using a thermistor temperature sensor. With Auto-Select selected, the EC100 will detect that the IRGASON is connected and will report ambient air temperature measurements from the thermistor. To diagnose problems with the temperature measurement, a fixed temperature value may be used, or the temperature sensor may be selected manually. 7.10 Fixed Temperature Value If the Temperature Sensor setting is None, the IRGASON will use the value of this setting for the sample temperature. This mode is intended for troubleshooting only. In normal operation, Temperature Sensor is set to Auto-Select, and this setting is not used. 17