Output Signals, 5.3 Physical Description, µV 16 bit
View all Campbell Scientific IRGASON manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 15 highlights
IRGASON® Integrated CO2/H2O Open-Path Gas Analyzer and 3D Sonic Anemometer 5.2 Output Signals Features • EC100 electronics outputs data using: o CS SDM o RS-485 o USB o Analog out Digital SDM (Synchronous Device for Measurement)1 Data type: FLOAT RS-485 Data type: Output Rate: Baud rate: ASCII 5 to 50 Hz 1200 to 230400 bps USB Data type: Output rate: ASCII 10, 25 or 50 Hz Analog (two outputs for CO2 and H2O densities) Voltage range: 0 mV to 5000 mV Resolution: 76 µV (16 bit) Update rate: 100 Hz Accuracy (at 25oC): ±3 mV CO2 density equation: Full scale range: mg·m-3 = 0.38632· (mVout) - 102.59 -103 to 1829 mg·m-3 H2O density equation: Full scale range: g·m-3 = 0.00865· (mVout) - 2.26 -2 to 41 g·m-3 1 Synchronous Device for Measurement. A Campbell Scientific, Inc. proprietary serial interface for datalogger to peripheral and sensor communication. See Section 8.1, SDM Output, for details. 2 user selectable 5.3 Physical Description Optical path length: 15.37 cm (6.05 in) Sonic path length: 10.0 cm Sonic path angle from vertical: 30° Spatial separation between gas analyzer and sonic volumes: 0 cm Dimensions IRGASON Length: Height: Cable Length: EC100 Enclosure: 51.5 cm (20.3 in) 31.1 cm (12.2 in) 3 m (9.8 ft) 24.1 x 35.6 x 14 cm (9.5 x 14.0 x 5.5 in) Weight IRGASON with Cable: EC100 with Enclosure: 2.8 kg (6.1 lbs) 3.2 kg (7 lbs) 5