Fluke Ti200 Manual - Page 51
Detailed Specifications
View all Fluke Ti200 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 51 highlights
Thermal Imagers Detailed Specifications Detailed Specifications Temperature Measurements Temperature Range (not calibrated below -10 °C) Ti200, Ti300 20 °C to +650 °C Ti400 20 °C to +1200 °C Accuracy 2 °C or 2 % (whichever is greater) at 25 °C ambient On-screen emissivity correction all models On-screen reflected background temperature compensation all models On-screen transmission correction all models Imaging Performance Image Capture Frequency 9 Hz or 60 Hz refresh rate depending on model variation Detector Type: Focal Plane Array, uncooled microbolometer Ti400 320 X 240 Ti300 240 X 180 Ti200 200 X 150 Thermal Sensitivity (NETD) Ti400, Ti300 0.05 °C at 30 °C target temp (50 mK) Ti200 0.075 °C at 30 °C target temp (75 mK) Total pixels Ti400 76,800 Ti300 43,200 Ti200 30,000 Infrared spectral band 7.5 μm to 14 μm (long wave) Visual (Visible Light) Camera Type Industrial performance 5.0 megapixel Minimum parallax alignment with standard IR lens 60 cm (~24 in) Standard Infrared Lens Field of View 24 ° x 17 ° Spatial Resolution (IFOV) Ti400 1.31 mRad Ti300 1.75 mRad Ti200 2.09 mRad Minimum Focus Distance 15 cm (6 in) MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.544152