Table 4-2. Volts DC Measurement Verification Points, Range, 5500A output V dc, Input A&B, Reading

Page 34 highlights

Fluke 19xB-19xC-2x5C Service Manual Table 4-2. Volts DC Measurement Verification Points Range 2 mV/div 1) 5 mV/div 10 mV/div 20 mV/div 50 mV/div 100 mV/div 200 mV/div 500 mV/div 1 V/div 2 V/div 5 V/div 10 V/div 20 V/div 50 V/div 100 V/div 1) C versions only. 5500A output V dc +6.0 mV -6.0 mV +15.0 mV -15.0 mV +30.0 mV -30.0 mV +60.0 mV -60.0 mV +150 mV -150 mV +300 mV -300 mV +600 mV -600 mV +1.50 V -1.50 V +3.00 V -3.00 V +6.00 V -6.00 V +15.0 V -15.0 V +30.0 V -30.0 V +60.0 V -60.0 V +150 V -150 V +300 V -300 V Input A&B Reading +4.9 to +7.1 -4.9 to -7.1 +14.3 to +15.7 -14.3 to -15.7 +29.1 to +30.9 -29.1 to -30.9 +58.6 to +61.4 -58.6 to -61.4 +143 to +157 -143 to -157 +291 to +309 -291 to -309 +586 to +614 -586 to -614 +1.43 to +1.57 -1.43 to -1.57 +2.91 to +3.09 -2.91 to -3.09 +5.86 to +6.14 -5.86 to -6.14 +14.3 to +15.7 -14.3 to -15.7 +29.1 to +30.9 -29.1 to -30.9 +58.6 to +61.4 -58.6 to -61.4 +143 to +157 -143 to -157 +291 to +309 -291 to -309 4-10

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86

Fluke 19xB-19xC-2x5C
Service Manual
4-10
Table 4-2. Volts DC Measurement Verification Points
Range
5500A output V dc
Input A&B
Reading
2 mV/div
1)
+6.0 mV
+4.9 to +7.1
-6.0 mV
-4.9 to -7.1
5 mV/div
+15.0 mV
+14.3 to +15.7
-15.0 mV
-14.3 to -15.7
10 mV/div
+30.0 mV
+29.1 to +30.9
-30.0 mV
-29.1 to -30.9
20 mV/div
+60.0 mV
+58.6 to +61.4
-60.0 mV
-58.6 to -61.4
50 mV/div
+150 mV
+143 to +157
-150 mV
-143 to -157
100 mV/div
+300 mV
+291 to +309
-300 mV
-291 to -309
200 mV/div
+600 mV
+586 to +614
-600 mV
-586 to -614
500 mV/div
+1.50 V
+1.43 to +1.57
-1.50 V
-1.43 to -1.57
1 V/div
+3.00 V
+2.91 to +3.09
-3.00 V
-2.91 to -3.09
2 V/div
+6.00 V
+5.86 to +6.14
-6.00 V
-5.86 to -6.14
5 V/div
+15.0 V
+14.3 to +15.7
-15.0 V
-14.3 to -15.7
10 V/div
+30.0 V
+29.1 to +30.9
-30.0 V
-29.1 to -30.9
20 V/div
+60.0 V
+58.6 to +61.4
-60.0 V
-58.6 to -61.4
50 V/div
+150 V
+143 to +157
-150 V
-143 to -157
100 V/div
+300 V
+291 to +309
-300 V
-291 to -309
1)