Table 4-2. Volts DC Measurement Verification Points, Range, 5500A output V dc, Input A&B, Reading
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 34 highlights
Fluke 19xB-19xC-2x5C Service Manual Table 4-2. Volts DC Measurement Verification Points Range 2 mV/div 1) 5 mV/div 10 mV/div 20 mV/div 50 mV/div 100 mV/div 200 mV/div 500 mV/div 1 V/div 2 V/div 5 V/div 10 V/div 20 V/div 50 V/div 100 V/div 1) C versions only. 5500A output V dc +6.0 mV -6.0 mV +15.0 mV -15.0 mV +30.0 mV -30.0 mV +60.0 mV -60.0 mV +150 mV -150 mV +300 mV -300 mV +600 mV -600 mV +1.50 V -1.50 V +3.00 V -3.00 V +6.00 V -6.00 V +15.0 V -15.0 V +30.0 V -30.0 V +60.0 V -60.0 V +150 V -150 V +300 V -300 V Input A&B Reading +4.9 to +7.1 -4.9 to -7.1 +14.3 to +15.7 -14.3 to -15.7 +29.1 to +30.9 -29.1 to -30.9 +58.6 to +61.4 -58.6 to -61.4 +143 to +157 -143 to -157 +291 to +309 -291 to -309 +586 to +614 -586 to -614 +1.43 to +1.57 -1.43 to -1.57 +2.91 to +3.09 -2.91 to -3.09 +5.86 to +6.14 -5.86 to -6.14 +14.3 to +15.7 -14.3 to -15.7 +29.1 to +30.9 -29.1 to -30.9 +58.6 to +61.4 -58.6 to -61.4 +143 to +157 -143 to -157 +291 to +309 -291 to -309 4-10