Fluke 225C Service Manual - Page 39
Input A&B Phase Measurements Test, READING 1, RECALL, SCREEN+SETUP, RECALL SETUP, Phase
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 39 highlights
4 Performance Verification 4.6 Scope Input A&B Tests • Press • Press READING 1 , and select on A | Hz. READING 2 , and select on B | Hz. 3. Using and select range 100 mV/div for A and B. 4. Using select the required time base setting. 5. Set the 5500A to source a sine wave according to the first test point in Table 4-6. As no 50Ω termination is applied, the 5500 leveled sine wave output amplitude will be twice the set value. 6. Observe the readings (1.A and 2.B) and check to see if it is within the range shown under the appropriate column. 7. Continue through the test points. 8. When you are finished, set the 5500A to 0 (zero) Volt, and to Standby. Table 4-6. Input A&B Frequency Measurement Accuracy Test Model Time base 5500A-SC... MODE Voltage all 20 ms/div wavegen, sine 600 mVpp 192B-C 20 ns/div levsine 300 mVpp 196B-C, 215C 20 ns/div levsine 300 mVpp 199B-C, 225C 20 ns/div levsine 300 mVpp Frequency 16 Hz 60 MHz 100 MHz 200 MHz Input A&B Reading 15.90 to 16.10 59.68 to 60.32 99.3 to 100.7 198.8 to 201.2 Note Duty Cycle and Pulse Width measurements are based on the same principles as Frequency measurements. Therefore the Duty Cycle and Pulse Width measurement function will not be verified separately. 4.6.7 Input A&B Phase Measurements Test Proceed as follows to test the phase measurement accuracy: 1. Connect the test tool to the 5500A as for the previous test (see Figure 4-4). 2. Select the following test tool setup: • Recall the created SETUP 1 (see section 4.4.3): press , select SCREEN+SETUP 1 , press RECALL SETUP . • Press • Press • Press READING 1 , and select on A | Phase. READING 2 , and select on B | Phase. 3. Using and select range 100 mV/div for A and B. RECALL , 4. Using select the required time base setting. 5. Set the 5500A to source a sine wave according to the first test point in Table 4-6. As no 50Ω termination is applied, the 5500 leveled sine wave output amplitude will be twice the set value. 4-15