Fluke 225C Service Manual - Page 84
Reassembling the Main PCA Unit
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 84 highlights
Fluke 19xB-19xC-2x5C Service Manual 5. To get access to the input circuits on the PCA, unscrew the Torx screws item 7 and remove the metal input circuit shielding boxes. Figure 6-7. PCA Unit Assembly ST8676.WMF 6.2.10 Reassembling the Main PCA Unit Reassembling the main PCA unit is the reverse of disassembly (see figure 6.7). However you must follow special precautions when reassembling the main PCA unit. 1. Install the metal input circuit shielding boxes (items 6) carefully. Take care that the notches at the edges of the boxes match the holes in the PCA. The plate spring in the Input A and Input B box must touch the C-ASIC N1000 (Input A) or N1200 (Input B) for cooling. Do not bend the springs! Caution A good thermal coupling between the C-ASIC's (N1000, N1200) and the input boxes is achieved by self adhesive thermal conductive pads. These pads can either be stuck on the spring in the box, or on the C-ASIC. If stuck on the C-ASIC, you can reuse the pad when replacing the C-ASIC. 2. Attach the isolation strip carefully! Insert the ends of the strip into the slots in the PCA, and push firmly until the strip is in its original position. 3. Put the PCA in the shielding box, and fasten the 2 hexagonal standoffs (item 10). 4. Attach the shielding cover (item 8). Ensure that the small optical gate PCA mounted on the main PCA sticks through the slot in the shielding cover. 5. Ensure that the rubber sealing ring (item 5) for the power connector is present 6-8