Fluke 225C Service Manual - Page 50
for PAL, PALplus or SECAM
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 50 highlights
Fluke 19xB-19xC-2x5C Service Manual 5. Using select test tool line number: ⇒ 310 for PAL, PALplus or SECAM ⇒ 262 for NTSC 6. Set the calibrator format and marker line number to : ⇒ PAL 310 (odd), for PAL and PALplus ⇒ SECAM 310 (odd), for SECAM ⇒ NTSC 262 odd, for NTSC. 7. Observe the trace, and check to see if the test tool triggers on the negative pulse before the marker. 8. Select the following test tool setup: • Press to open the Trigger Options menu. • Choose VIDEO on A... , then from the shown opened menu choose Polarity: NEGATIVE | PAL ( or NTSC PALplus SECAM ) 9. Set the calibrator video trigger output signal to -100% 10. Using select line number 310 (PAL, PALplus or SECAM) or 262 (NTSC) 11. Set the calibrator format and marker line number to : ⇒ PAL 310 (odd), for PAL and PALplus ⇒ SECAM 310 (odd), for SECAM ⇒ NTSC 262 odd, for NTSC. 12. Observe the trace, and check to see if the test tool triggers on the positive pulse before the marker. Figure 4-17. SC600 Marker Pulse video-sc600.bmp 4-26