Fluke 225C Service Manual - Page 48
Video test using SC600 Scope Calibration Option, Polarity, NEGATIVE, PALplus, SECAM
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 48 highlights
Fluke 19xB-19xC-2x5C Service Manual Figure 4-13. Test Tool Input A to TV Signal Generator Inverted al-tv-ai.bmp 8. Select the following test tool setup: • Press to open the Trigger Options menu. • Choose VIDEO on A... , then from the shown opened menu choose Polarity: NEGATIVE | PAL ( or NTSC PALplus SECAM ) 9. Using select line number 310 (PAL or SECAM) or 262 (NTSC) 10. Observe the trace, and check to see if the test tool triggers on line number 310 (PAL or SECAM, see Figure 4-14), or line number 262 (NTSC, see Figure 4-15). Figure 4-14. Trace for PAL/SECAM line 310 Negative Video Figure 4-15. Trace for NTSC line 262 Negative Video 4.6.14 Video test using SC600 Scope Calibration Option You can skip this test if you did test 4.6.13 Video test using the Video Pattern Generator. Only one of the systems NTSC, PAL, PALplus, or SECAM has to be verified. 4-24