Fluke 225C Service Manual - Page 36
Input A & B AC Coupled Lower Frequency, RECALL, SCREEN+SETUP, RECALL SETUP, READING 1
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 36 highlights
Fluke 19xB-19xC-2x5C Service Manual Table 4-3. Volts AC Measurement Verification Points Range 2 mV/div 1) (Select 10 ms/div) Set input A&B Bandwidth 10 kHz to prevent OL due to calibrator noise: see step 2. 5 mV/div (Select 20 μs/div). Set input A&B Bandwidth 20 MHz 10 mV/div 20 mV/div 50 mV/div 100 mV/div 200 mV/div 500 mV/div (Select 10 ms/div) 500 mV/div (Select 20 μs/div) 1 V/div 2 V/div 5 V/div 10 V/div 20 V/div 50 V/div 100 V/div 1) C versions only 5500A output V ac Frequency 4 mV 60 Hz 10 mV 20 mV 40 mV 100 mV 200 mV 400 mV 900 mV 900 mV 2 V 4 V 9 V 20 V 40 V 90 V 200 V 20 kHz 20 kHz 20 kHz 20 kHz 20 kHz 20 kHz 60 Hz 20 kHz 20 kHz 20 kHz 20 kHz 20 kHz 20 kHz 20 kHz 20 kHz Input A&B Reading 3.0 mV to 5.0 mV 8.3 mV to 11.7 mV 18.0 mV to 22.0 mV 37.5 mV to 42.5 mV 96.0 mV to 104.0 mV 180 mV to 220 mV 375 mV to 425 mV 877 mV to 923 mV 863 mV to 937 mV 1.80 V to 2.20 V 3.75 V to 4.25 V 8.63 V to 9.37 V 18.0 V to 22.0 V 37.5 V to 42.5 V 86.3 V to 93.7 V 180 V to 220 V 4.6.4 Input A & B AC Coupled Lower Frequency Test Proceed as follows to test the ac coupled input low frequency accuracy: 1. Connect the test tool to the 5500A as for the previous test (see Figure 4-3). 2. Select the following test tool setup: • Recall the created SETUP 1 (see section 4.4.3): press , RECALL , select SCREEN+SETUP 1 , press RECALL SETUP . • Press • Press • Press READING 1 , and select on A | V ac. READING 2 , and select on B | V ac. • Press , then using select COUPLING AC • Press , then using select COUPLING AC 4-12