Fluke 225C Service Manual - Page 74
Possible error messages., WarmingUp CL 0200:IDLE
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 74 highlights
Fluke 19xB-19xC-2x5C Service Manual Possible error messages. The following messages can be shown on the test tool display: WARNING: Calibration data not valid. Save data and exit maintenance mode? Proceed as follows: • If you did the WarmingUp and Pre-Calibration successfully (section 5.5), and you want to store the Pre-Calibration data before continuing with the Final Calibration: ⇒ Press YES. When turning the test tool off and on again, it will show the message: The instrument needs calibration. Please contact your service center. The calibration date and number will not be updated. You must continue with the Final Calibration! • To return to the Maintenance mode, if you want to repeat the complete calibration: ⇒ Press NO. Now press until the display shows WarmingUp (CL 0200):IDLE, and calibrate the test tool, starting at section 5.5. • If you want to exit and maintain the old calibration data: ⇒ Turn the test tool off. 5-18