Fluke 225C Service Manual - Page 66
Continue at A SCOPE Output to Test Tool Input B
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 66 highlights
Fluke 19xB-19xC-2x5C Service Manual 5-10 Figure 5-4. 5500A SCOPE Output to Test Tool Input B al55scb.bmp 3. Set the 5500A SCOPE output to source the signal required for the first calibration point in Table 5-2. 4. Set the 5500A in operate (OPR) or standby (STBY) as indicated. 5. Press to start the calibration. 6. Wait until the display shows calibration status :READY . 7. Press to select the next calibration step, set the 5500A to the next calibration point signal, and start the calibration. Continue through all calibration points of Table 5-2. 8. When you are finished, set the 5500A to Standby. 9. Continue at Section 5.6.3. Cal step CL 0674 CL 0410 CL 0724 CL 0421 CL 0482 CL 0483 Table 5-2. Input B LF-HF Gain Calibration Points UUT input signal none 0.5 Vpp square wave, 1 kHz none 0.5 Vpp square wave, 1 kHz 0.5 Vpp sine wave, 50 kHz 0.5 Vpp sine wave Fluke 199B-C, 225C: 221 MHz Fluke 196B-C, 215C: 141 MHz Fluke 192B-C: 91 MHz 5500A Setting STANDBY SCOPE edge, 0.5 Vpp, 1 kHz STANDBY SCOPE edge, 0.5 Vpp, 1 kHz SCOPE levsine, 0.5 Vpp, 50 kHz SCOPE levsine, 0.5 Vpp, 221 MHz 141 MHz 91 MHz