Fluke 19xB-19xC-2x5C Service Manual 4.9 Probe Calibration Generator Test To verify the internal probe calibration square wave generator, you can do a Probe Calibration as described in section 5.8. If no square wave appears on the screen, either • the probe is defective: try another probe, check the probe with an external voltage in a scope application, or • the internal square wave generator is defective. This is the end of the Performance Verification Procedure. 4-32
Fluke 19xB-19xC-2x5C
Service Manual
4-32
4.9 Probe Calibration Generator Test
To verify the internal probe calibration square wave generator, you can do a Probe
Calibration as described in section 5.8.
If no square wave appears on the screen, either
•
the probe is defective: try another probe, check the probe with an external voltage in a
scope application,
or
•
the internal square wave generator is defective.
This is the end of the Performance Verification Procedure.