Fluke 225C Service Manual - Page 47
Trace for PAL/SECAM line 622, for PAL or SECAM
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 47 highlights
4 Performance Verification 4.6 Scope Input A&B Tests • sync pulse amplitude > 0.7 div. • chroma amplitude zero. 4. Observe the trace, and check to see if the test tool triggers on line number: ⇒ 622 for PAL or SECAM, see Figure 4-9 ⇒ 525 for NTSC, see Figure 4-10. Figure 4-9. Trace for PAL/SECAM line 622 Figure 4-10. Trace for NTSC line 525 5. Using select line number: ⇒ 310 for PAL or SECAM ⇒ 262 for NTSC 6. Observe the trace, and check to see if the test tool triggers on: ⇒ line number 310 for PAL or SECAM, see Figure 4-11. ⇒ line number 262 for NTSC, see Figure 4-12. Figure 4-11. Trace for PAL/SECAM line 310 Figure 4-12. Trace for NTSC line 262 7. Apply the inverted TV Signal Generator signal to the test tool. Invert the signal by using a Banana Plug to BNC adapter (Fluke PM9081/001) and a Banana Jack to BNC adapter (Fluke PM9082/001), as shown in Figure 4-13. 4-23