4 Performance Verification 4.6 Scope Input A&B Tests Proceed as follows: 1. Connect the test tool to the calibrator as shown in Figure 4-16. Figure 4-16. Test Tool Input A to TV Signal Generator al55sca.bmp 2. Select the following test tool setup: • Reset the test tool • Press , then press to open the Trigger Options menu. • Choose VIDEO on A... , then from the shown opened menu choose Polarity: POSITIVE | PAL ( or NTSC PALplus SECAM ) • Press to select ALL LINES • Press to enable the arrow keys for selecting the video line number. • Using select line number: ⇒ 622 for PAL, PALplus, or SECAM ⇒ 525 for NTSC. • Using and set the Input A sensitivity to 2 V/div (the actual probe setting is 10:1). • Using select the time base to 20 μs/div. 3. Set the calibrator to mode video with amplitude +100%. Set format and marker line number to : ⇒ PAL 622 (even), for PAL and PALplus ⇒ SECAM 622 (even), for SECAM ⇒ NTSC 262 even, for NTSC. 4. Observe the trace, and check to see if the test tool triggers on the negative pulse before the marker pulse (see Figure 17). 4-25

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86

Performance Verification
4.6 Scope Input A&B Tests
4
4-25
Proceed as follows:
1.
Connect the test tool to the calibrator as shown in Figure 4-16.
al55sca.bmp
Figure 4-16. Test Tool Input A to TV Signal Generator
2.
Select the following test tool setup:
Reset the test tool
Press
, then press
to open the Trigger Options menu.
Choose
VIDEO on A...
, then from the shown opened menu choose
Polarity:
POSITIVE |
PAL
(
or
NTSC
PALplus
SECAM
)
Press
to select
ALL LINES
Press
to enable the arrow keys for selecting the video line number.
Using
select line number:
622 for PAL, PALplus, or SECAM
525 for NTSC.
Using
and
set the Input A sensitivity to 2 V/div (the actual probe setting
is 10:1).
Using
select the time base to 20
μ
s/div.
3.
Set the calibrator to mode video with amplitude +100%.
Set format and
marker line
number to :
PAL 622 (even), for PAL and
PALplus
SECAM 622 (even), for SECAM
NTSC 262 even, for NTSC.
4.
Observe the trace, and check to see if the test tool triggers on the negative pulse
before the marker pulse (see Figure 17).