Page 49 highlights
4 Performance Verification 4.6 Scope Input A&B Tests Proceed as follows: 1. Connect the test tool to the calibrator as shown in Figure 4-16. Figure 4-16. Test Tool Input A to TV Signal Generator al55sca.bmp 2. Select the following test tool setup: • Reset the test tool • Press , then press to open the Trigger Options menu. • Choose VIDEO on A... , then from the shown opened menu choose Polarity: POSITIVE | PAL ( or NTSC PALplus SECAM ) • Press to select ALL LINES • Press to enable the arrow keys for selecting the video line number. • Using select line number: ⇒ 622 for PAL, PALplus, or SECAM ⇒ 525 for NTSC. • Using and set the Input A sensitivity to 2 V/div (the actual probe setting is 10:1). • Using select the time base to 20 μs/div. 3. Set the calibrator to mode video with amplitude +100%. Set format and marker line number to : ⇒ PAL 622 (even), for PAL and PALplus ⇒ SECAM 622 (even), for SECAM ⇒ NTSC 262 even, for NTSC. 4. Observe the trace, and check to see if the test tool triggers on the negative pulse before the marker pulse (see Figure 17). 4-25