Fluke 225C Service Manual - Page 69
Input A&B Volt Gain
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 69 highlights
5 Calibration Adjustment 5.6 Final Calibration 5.6.5 Input A&B Volt Gain Warning Dangerous voltages will be present on the calibration source and connection cables during the following steps. Ensure that the calibrator is in standby mode before making any connection between the calibrator and the test tool. Proceed as follows to do the Input A&B Volt Gain calibration. 1. Press to select the first calibration step in Table 5-4. 2. Connect the test tool to the 5500A as shown in Figure 5-6. Figure 5-6. Test tool Input A&B to 5500 Normal Output al55ab.bmp 3. Set the 5500A to supply a DC voltage (NORMAL output), to the first calibration point in Table 5-4. 4. Set the 5500A to operate (OPR). 5. Press to start the calibration. 6. Wait until the display shows calibration status :READY. 7. Press to select the next calibration step, set the 5500A to the next calibration point, and start the calibration. Continue through all calibration points of Table 5-4. 8. Set the 5500A to Standby, and continue at Section 5.6.6. 5-13