Fluke 225C Service Manual - Page 44

Input B Trigger Sensitivity Test, under the appropriate column of Table 4-10.

Page 44 highlights

Fluke 19xB-19xC-2x5C Service Manual 4.6.11 Input B Trigger Sensitivity Test Proceed as follows to test the Input B trigger sensitivity: 1. Connect the test tool to the 5500A as shown in Figure 4-7. Figure 4-7. 5500A Scope Output to Test Tool Input B al55scb.bmp 2. Select the following test tool setup: • Reset the test tool • Press and use to turn Input B on. • Press and use to turn Input A off. • Using move the Input B trace zero to the center grid line. • Press and use to select Input B as trigger source. • Using and change the sensitivity range to select manual sensitivity ranging, and lock the Input B sensitivity range on 2 V/div. 3. Using select the time base indicated under the first column of Table 4-10. 4. Set the 5500A to source the leveled sine wave given in the first row of Table 4-10. 5. Adjust the 5500A output voltage until the displayed trace has the amplitude indicated under the appropriate column of Table 4-10. 6. Verify that the signal is well triggered. If it is not, press , then using enable the up/down arrow keys for manual Trigger Level adjustment. Adjust the trigger level and verify that the signal will be triggered now. The trigger level is indicated by the trigger icon ( ). 7. Continue through the test points. 8. When you are finished, set the 5500A to Standby. 4-20

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86

Fluke 19xB-19xC-2x5C
Service Manual
4-20
4.6.11 Input B Trigger Sensitivity Test
Proceed as follows to test the Input B trigger sensitivity:
1.
Connect the test tool to the 5500A as shown in Figure 4-7.
al55scb.bmp
Figure 4-7. 5500A Scope Output to Test Tool Input B
2.
Select the following test tool setup:
Reset the test tool
Press
and use
to turn Input B on.
Press
and use
to turn Input A off.
Using
move the Input B trace zero to the center grid line.
Press
and use
to select Input B as trigger source.
Using
and
change the sensitivity range to select manual sensitivity
ranging, and lock the Input B sensitivity range on 2 V/div.
3.
Using
select the time base indicated under the first column of Table 4-10.
4.
Set the 5500A to source the leveled sine wave given in the first row of Table 4-10.
5.
Adjust the 5500A output voltage until the displayed trace has the amplitude indicated
under the appropriate column of Table 4-10.
6.
Verify that the signal is well triggered.
If it is not, press
, then using
enable the up/down arrow keys for manual
Trigger Level adjustment.
Adjust the trigger level and verify that the signal will be
triggered now.
The trigger level is indicated by the trigger icon (
).
7.
Continue through the test points.
8.
When you are finished, set the 5500A to Standby.