Fluke 225C Service Manual - Page 62
Contrast Calibration Adjustment, WarmingUp CL 0200:IDLE valid, Contrast CL 0100, CALIBRATE
View all Fluke 225C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 62 highlights
Fluke 19xB-19xC-2x5C Service Manual 5.4 Contrast Calibration Adjustment After entering the calibration mode the display shows: WarmingUp (CL 0200):IDLE (valid) Do not press now! If you did, turn the test tool off and on, and enter the calibration mode again. Proceed as follows to adjust the maximum display darkness (CL 0100), the default contrast (CL 0110) , and the maximum display brightness (CL 0120). 1. Press three times to select maximum darkness calibration Contrast (CL 0100): 2. Press CALIBRATE . The display will show a dark test pattern, see Figure 5-2 3. Using adjust the display to the maximum darkness at which the test pattern is only just visible. 4. Press to return to the softkey menu. 5. Press 6. Press 7. Using to select default contrast calibration Contrast (CL 0110): CALIBRATE. The display shows the test pattern at default contrast. set the display to optimal (becomes default) contrast. 8. Press to return to the softkey menu. 9. Press to select maximum brightness calibration Contrast (CL 0120): 10. Press CALIBRATE. The display shows a bright test pattern. 11. Using adjust the display to the maximum brightness, at which the test pattern is only just visible. 12. Press to return to the softkey menu. 13. Now you can either • Exit, if only the Contrast had to be adjusted. Continue at Section 5.7. or • Do the complete calibration. Press and continue at Section 5.5. to select the next step (WarmingUp), Figure 5-2. Display Test Pattern 5-6