Fluke 8808A User Manual - Page 19

Electrical, Remote Interfaces, Warranty, Electrical Specifications, DC Voltage Specifications

Page 19 highlights

1 Introduction and Specifications Electrical Specifications Electrical Input Protection 1000 V all ranges Overrange 10 % on the largest ranges of all functions except continuity and diode test Remote Interfaces RS-232C Warranty One year Electrical Specifications Accuracy specifications are valid for 5-½ digit mode and after at least a half-hour warm-up DC Voltage Specifications Maximum Input 1000 V on any range Common Mode Rejection 120 dB at 50 or 60 Hz ±0.1% (1 kΩ unbalance) Normal Mode Rejection 80 dB at Slow Rate A/D Nonlinearity 15 ppm of range Input Bias Current 30 pA at 25 °C Settling Considerations Measurement settling times are affected by source impedance, cable dielectric characteristics, and input signal changes Input Characteristics Range Full-Scale (5-1/2 Digits) Slow Resolution Medium Fast Input Impedance 200 mV 2 V 20 V 200 V 1000 V 199.999 mV 1.99999 V 19.9999 V 199.999 V 1000.00 V 1 μV 10 μV 100 μV 1 mV 10 mV 10 μV 100 μV 1000 μV 10 mV 100 mV 10 μV 100 μV 1000 μV 10 mV 100 mV >10 GΩ[1] >10 GΩ[1] 10 MΩ±1 % 10 MΩ±1 % 10 MΩ±1 % Notes: [1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 MΩ. Accuracy Accuracy [1] Range 90 days 23 °C ± 5 °C 200 mV 0.01 + 0.003 2 V 0.01 + 0.002 20 V 0.01 + 0.003 200 V 0.01 + 0.002 1000 V 0.01 + 0.002 Notes: [1] Accuracy given as ± (% of reading + % of range) 1 year 23 °C ± 5°C 0.015 + 0.004 0.015 + 0.003 0.015 + 0.004 0.015 + 0.003 0.015 + 0.003 Temperature Coefficient/°C Outside 18 - 28 °C 0.0015 + 0.0005 0.001 + 0.0005 0.0020 + 0.0005 0.0015 + 0.0005 0.0015 + 0.0005 1-9

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100

Introduction and Specifications
Electrical Specifications
1
1-9
Electrical
Input Protection
...............................................
1000 V all ranges
Overrange
..........................................................
10 % on the largest ranges of all functions except continuity and diode
test
Remote Interfaces
RS-232C
Warranty
One year
Electrical Specifications
Accuracy specifications are valid for 5-½ digit mode and after at least a half-hour warm-up
DC Voltage Specifications
Maximum Input
.................................................
1000 V on any range
Common Mode Rejection
.................................
120 dB at 50 or 60 Hz
±
0.1
(1 k
Ω
unbalance)
Normal Mode Rejection
....................................
80 dB at Slow Rate
A/D Nonlinearity
................................................
15 ppm of range
Input Bias Current
............................................
<30 pA at 25
°
C
Settling Considerations
...................................
Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes
Input Characteristics
Resolution
Range
Full-Scale
(5-1/2 Digits)
Slow
Medium
Fast
Input Impedance
200 mV
199.999 mV
1
μ
V
10
μ
V
10
μ
V
>10 G
[1]
2 V
1.99999 V
10
μ
V
100
μ
V
100
μ
V
>10 G
[1]
20 V
19.9999 V
100
μ
V
1000
μ
V
1000
μ
V
10 M
±1 %
200 V
199.999 V
1 mV
10 mV
10 mV
10 M
±1 %
1000 V
1000.00 V
10 mV
100 mV
100 mV
10 M
±1 %
Notes:
[1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M
.
Accuracy
Accuracy
[1]
90 days
1 year
Range
23 °C ± 5 °C
23 °C ± 5°C
Temperature Coefficient/°C
Outside 18 – 28 °C
200 mV
0.01 + 0.003
0.015 + 0.004
0.0015 + 0.0005
2 V
0.01 + 0.002
0.015 + 0.003
0.001 + 0.0005
20 V
0.01 + 0.003
0.015 + 0.004
0.0020 + 0.0005
200 V
0.01 + 0.002
0.015 + 0.003
0.0015 + 0.0005
1000 V
0.01 + 0.002
0.015 + 0.003
0.0015 + 0.0005
Notes:
[1]
Accuracy given as ± (% of reading + % of range)