4 Operating the Meter Using the Computer Interface Status Registers The ESE is a mask register that allows the host to enable or disable (mask) each bit in the ESR. When a bit in the ESE is set to 1, the corresponding bit in the ESR is enabled. When any enabled bit in the ESR changes from 0 to 1, the ESB bit in the STB also changes to 1. When the ESR is read using the *ESR? command or cleared using the *CLS command, the ESB bit in the STB returns to 0. Power On User Request Command Error Execution Error Device Dependent Error Query Error Request Control Operation Complete Logical OR 7 65432 10 & & & && & && Standard Event Status Register *ESR? Standard 7 6 5 4 3 2 1 0 Event Status Enable Register *ESE *ESE? Summary Message Event Summary Bit (ESB) (Bit 5 of Status Byte Register) Figure 4-3. Event Status and Event Status Enable Registers aam22f.eps 4-13