Fluke 8808A User Manual - Page 82
Table 4-11. Range and Measurement Rate Commands and Queries cont.
View all Fluke 8808A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 82 highlights
8808A Users Manual Table 4-11. Range and Measurement Rate Commands and Queries (cont.) Command RANGE Description Sets the primary display to where is the number in the Range Value column of Table 4-11A that corresponds with the applicable function ranges (voltage, ohms, current, etc.). RANGE1? RANGE2? RATE RATE? Table 4-11A. Ranges for Each Function Range Value Voltage Range Ohms Range AC Current Freq. Range DC Current 1 200 mV 200 e 20 mA 2 kHz 200 µA 2 2V 2 ke 200 mA 20 kHz 2000 µA 3 20 V 20 ke 2 A 200 kHz 20 mA 4 200 V 200 ke 5 1000 V dc[1] 2 Me 10 A NA 1000 kHz 200 mA NA 2 A 6 NA 20 Me NA NA 10 A 7 NA 100 Me NA NA NA [1] 1000 V dc, 750 V ac Returns the range presently selected on the primary display. Returns the range presently selected on the secondary display. If the secondary display is inactive, an Execution Error is generated. Sets the measurement rate to where is either S for slow (2.5 readings/second), M for medium (20 readings/second), or F for fast (100 readings/second). S, M, and F can be entered as uppercase or lowercase. Any other entry for generates an Execution Error. Returns as S for slow (2.5 readings/second), M for medium (20 readings/second), or F for fast (100 readings/second). 4-22