Fluke 8808A User Manual - Page 48
Automatic Input Terminal Detection, Current Measurement <200 mA
View all Fluke 8808A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 48 highlights
8808A Users Manual I < 200 mA V AC 2W/4W HI INPUT V SENSE 4W HI 1000V 750V MAX LO 200 mA MAX 1000 V 600 V CAT I CAT II 300 V 1V LO 500 Vpk 10 A MAX mA 10 A Figure 3-8. Current Measurement 200 mA and I < 10 A V AC 2W/4W HI INPUT V SENSE 4W HI 1000V 750V MAX LO 200 mA MAX 1000 V 600 V CAT I CAT II 300 V 1V LO 500 Vpk 10 A MAX mA 10 A Figure 3-9. Current Measurement 200 mA to 10 A eue14.eps Automatic Input Terminal Detection For ac and dc current measurement functions, the Meter automatically detects the signal input between the mA and 10 A input terminals. A front-panel annunciator indicates whether the Meter is in the mA or A ranges. If a test lead is inserted in to the mA input terminal and there isn't a test lead in the 10 A terminal, then only the 200 µA to 200 mA ranges are selectable. If the 10 A terminal has a test lead, then only the 2 A and 10 A ranges are selectable. 3-14