Fluke 8808A User Manual - Page 85
Miscellaneous Commands and Queries, RS-232 Remote / Local Configurations
View all Fluke 8808A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 85 highlights
4 Operating the Meter Using the Computer Interface Computer Interface Command Set Miscellaneous Commands and Queries Table 4-15 describes miscellaneous commands and queries. Command ^C (CONTRL C) FORMAT FORMAT? PRINT SERIAL? Table 4-15. Miscellaneous Commands and Queries Description Causes => to be output. Set output to 1 or 2. Format 1 outputs measurement values without measurement units (VDC, ADC, OHMS, etc.). Format 2 allows measurement units to be output with measurement units. (See Table 4-16.) Format 2 primarily is used with RS-232 print-only mode. Returns the format in use (1 or 2). Sets print rate for print mode. See Table 4-2. Returns Meter's serial number. Table 4-16. Measurement Units Output with Format 2 Measurement Function Units Output Volts dc Volts ac Amps dc Amps ac Resistance Frequency DIODE Continuity Test VDC VAC ADC AAC OHMS HZ VDC OHMS RS-232 Remote / Local Configurations Table 4-17 describes the RS-232 remote and local configuration commands, which are used with the RS-232 interface to set up the remote/local configuration of the Meter. These commands are valid only when the RS-232 interface is enabled. 4-25