Fluke 8808A User Manual - Page 22
AC Current, Accuracy, Input Characteristics
View all Fluke 8808A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 22 highlights
8808A Users Manual Accuracy Accuracy [1] Range 90 days 23 °C ± 5 °C 200 μA 2 mA 20 mA 200 mA 2 A 10 A 0.02 + 0.005 0.015 + 0.005 0.03 + 0.02 0.02 + 0.005 0.05 + 0.02 0.18 + 0.01 Notes: [1] Accuracy given as ± (% of reading + % of range) 1 year 23 °C ± 5 °C 0.03 + 0.005 0.02 + 0.005 0.04 + 0.02 0.03 + 0.008 0.08 + 0.02 0.2 + 0.01 Temperature Coefficient/°C Outside 18 - 28 °C 0.003 + 0.001 0.002 + 0.001 0.005 + 0.001 0.005 + 0.001 0.008 + 0.001 0.008 + 0.001 AC Current The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs from 1 % to 5 % of range, add an additional error of 0.1 % of range. Input Protection Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses Measurement Method AC-coupled True RMS Shunt Resistance 0.01 Ω for 2 A and 10 A ranges 1 Ω for 20 mA and 200 mA AC Filter Bandwidth: Slow 20 Hz - 100 kHz Maximum Crest Factor 3:1 at Full Scale Additional Crest Factor Errors (