Fluke 8808A User Manual - Page 45
Frequency Ranging, Measuring Resistance, 2-Wire Resistance Measurement
![]() |
View all Fluke 8808A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 45 highlights
3 Operating the Meter from the Front Panel Selecting a Measurement Function Frequency Ranging Frequency measurements are automatically ranged so that a frequency measurement is always displayed with maximum resolution. To select a range manually, press F to select the frequency function, and then press U or V to select a range manually. Manual ranging can be performed on readings displayed in the primary display only. If you manually select a frequency range and the measurement exceeds the full-scale value of that range, 0L is displayed to indicate an overload. If the frequency is lower than 20 Hz, UL is displayed. Refer to the "Electrical Specifications" section in Chapter 1 for frequency ranges and full-scale values. Measuring Resistance The Meter offers 2-wire and 4-wire ohms measurement. Press O to toggle between 2wire and 4-wire measurement modes. The Meter displays 2*4 Wire at 2 wire or 2x4 wire resistance measurement and displays 4 Wire at 4 wire resistance measurement. 2-Wire Resistance Measurement To perform a 2-wire resistance measurement: 1. Connect test leads between the Meter and the circuit under test as shown in Figure 3-5. 2. If required, press O to select 2-wire resistance measurement mode. 2*4 Wire is displayed. The function and measurement are displayed. 2W/4W HI INPUT V SENSE 4W HI 1000V 750V MAX LO 200 mA MAX 1000 V 600 V CAT I CAT II 300 V 1V LO 500 Vpk 10 A MAX mA 10 A Figure 3-5. 2-Wire Resistance Measurement eue11.eps 3-11