Invacare M61 Owners Manual - Page 27
Service Indicator Light Diagnostics - diagnostic codes
View all Invacare M61 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 27 highlights
SECTION 4-SAFETY INSPECTION/TROUBLESHOOTING DISPLAY Information Gauge Display DESCRIPTION DEFINITION COMMENTS Left RED LED is flashing. Battery charge is low. The batteries should be charged as soon as possible. Left to Right "chase" alter- Joystick is in program- The steady LEDs indicate nating with steady display. ming, inhibit and/or charg- the current state of the ing mode. battery charge. All LEDs are flashing slowly. Joystick has detected Out- Release the joystick back of-Neutral-at-Power-Up to Neutral. mode. Service Indicator Light Diagnostics NUMBER OF FLASHES 1 2 3 4 5 6 7 8 9 10 11 ERROR CODE DESCRIPTION User Fault Battery Fault Left Motor Fault Right Motor Fault Left Park Brake Fault Right Park Brake Fault Remote Fault Controller Fault Communications Fault General Fault Incompatible or incorrect Remote POSSIBLE SOLUTION Release joystick to neutral and try again. Charge the batteries. Check that battery cables are connected properly. If necessary, replace batteries. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Check to make sure joystick is connected properly. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Contact Invacare/Dealer for service. Wrong type of remote connected. Contact Invacare/Dealer for service. Part No 1125085 27 Pronto® M51™and M61™with SureStep®