Campbell Scientific CR10 CR10 Measurement and Control - Page 168

Campbell Scientific CR10 Manual

Page 168 highlights

SECTION 13. CRlO MEASUREMENTS Setup Amplifier In tegrotion Settlin g A/D Conversion 450 uS 25OuS fost 70OuS 2.72mS slow Reset Integrotor F|GURE 13.2-1. Timing of Singld:Ended Measurement 13.2 SINGLE.ENDED AND DIFFERENTIAL VOLTAGE MEASUREMENTS NOTE: The channel numbering on the old silver CR10 wiring panel refers to ditferential channels. Either the high or low side of a differentialchannel can be used for single-ended measurements. Each side must be counted when numbering singleended channels; e.9., the high and low sides of difterentialchannel4 are singleended channels 7 and 8, respectively. The timing and sequence of a single-ended measurement is shown in Figure 13.2-1. A single-ended measurement is made on a single input which is referenced to ground. A single integration is pedormed for each measurement. A differential measurement measures the difference in voltage between two inputs. The measurement sequence on a differential measurement involves two integrations. First with the high input referenced to the low, then with the inputs reversed (Figure 13.2-2). The CR10 computes the differentialvoltage by averaging the magnitude of the results from the two integrations and using the polarity from the first. An exception to this is the differential measurement in Instruction 8 which makes only one integration. Because a single-ended measurement is referenced to CR10 ground, any difference in ground potential between the sensor and the CR10 will result in an error in the measurement. For example, if the measuring junction of a copper-constantan thermocouple, used to measure soiltemperature, is not insulated and the potential of earth ground is 1 mV greater at the sensor than at the point where the CR10 is grounded, the measured voltage would be 1 greater than the thermocouple output or approximately 25'C high. Another instance where a ground potential difference creates a problem is in a case such as described in Section 7.2, where external signalconditioning circuitry is powered from the same source as the CR10. Despite being tied to the same ground, differences in current drain and lead resistance result in different ground potential at the two instruments. For this a differential measurement should be made on an analog output from the external signal conditioner. Differential measurements MUST be used where the inputs are known to be different from ground, such as is the case with the output from a full bridge. Setup & Integ. A/D Conv. Switch lnpt. lnteg. A/D Conv. Settle & Settle 250us fost 700us 600us 250us fost 700us 2.72ms slow 2.72ms slow FIGURE 13.2-2. Ditferential Voltage Measurement Sequence 13-2

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159
  • 160
  • 161
  • 162
  • 163
  • 164
  • 165
  • 166
  • 167
  • 168
  • 169
  • 170
  • 171
  • 172
  • 173
  • 174
  • 175
  • 176
  • 177
  • 178
  • 179
  • 180
  • 181
  • 182
  • 183
  • 184
  • 185
  • 186
  • 187
  • 188
  • 189
  • 190
  • 191
  • 192
  • 193
  • 194
  • 195
  • 196
  • 197
  • 198
  • 199
  • 200
  • 201
  • 202
  • 203
  • 204
  • 205
  • 206
  • 207
  • 208
  • 209
  • 210
  • 211
  • 212
  • 213
  • 214
  • 215
  • 216
  • 217
  • 218
  • 219
  • 220
  • 221
  • 222
  • 223
  • 224
  • 225
  • 226
  • 227
  • 228
  • 229
  • 230
  • 231
  • 232
  • 233
  • 234
  • 235
  • 236
  • 237
  • 238

SECTION
13.
CRlO
MEASUREMENTS
Setup
Amplifier
Settlin
g
In
tegrotion
A/D
Conversion
450
uS
13.2
SINGLE.ENDED AND
DIFFERENTIAL VOLTAGE
MEASUREMENTS
Setup
&
Integ.
Settle
25OuS
fost
70OuS
2.72mS
slow
Reset Integrotor
Because a single-ended measurement
is
referenced
to
CR10 ground, any
difference
in
ground potential between the sensor and the
CR10
will
result in an
error
in
the
measurement.
For example,
if
the
measuring
junction
of a
copper-constantan
thermocouple,
used
to
measure
soiltemperature,
is not insulated and
the
potential
of
earth ground
is
1
mV greater at
the sensor
than
at
the
point
where
the
CR10
is
grounded,
the
measured voltage
would be
1
greater than the thermocouple output or
approximately
25'C
high.
Another instance where a ground potential
difference creates a problem is
in a
case such
as described in Section 7.2, where external
signalconditioning
circuitry is powered from the
same source as the
CR10.
Despite being tied
to
the
same ground,
differences
in
current drain
and lead resistance result
in
different ground
potential
at
the
two instruments.
For
this
a
differential
measurement should be made
on
an analog output from
the
external signal
conditioner.
Differential measurements MUST
be used where
the
inputs
are
known
to
be
different
from
ground, such
as
is
the case with
the output
from
a
full
bridge.
F|GURE 13.2-1.
Timing of Singld:Ended
Measurement
NOTE:
The
channel
numbering on the old
silver CR10
wiring
panel refers
to
ditferential
channels.
Either
the
high
or
low
side of
a
differentialchannel can
be used
for
single-ended measurements.
Each side
must
be
counted
when
numbering single-
ended channels; e.9.,
the
high and low
sides of
difterentialchannel4
are single-
ended channels 7 and
8,
respectively.
The
timing
and sequence of a single-ended
measurement is
shown
in
Figure 13.2-1. A
single-ended measurement is made on a single
input which is referenced
to
ground.
A single
integration is
pedormed
for
each
measurement.
A
differential
measurement measures
the
difference
in
voltage
between
two
inputs.
The
measurement sequence on a differential
measurement involves
two integrations.
First
with
the
high input referenced to
the
low,
then
with
the
inputs reversed
(Figure
13.2-2).
The CR10 computes the
differentialvoltage
by
averaging
the
magnitude of
the
results from
the
two
integrations and using
the
polarity from
the
first.
An exception to
this
is
the differential
measurement
in
Instruction
8
which
makes only
one integration.
A/D
Conv.
Switch
lnpt.
&
Settle
lnteg.
A/D
Conv.
250us
fost
700us
2.72ms
slow
600us
250us
fost
700us
2.72ms
slow
13-2
FIGURE
13.2-2. Ditferential Voltage
Measurement Sequence