Campbell Scientific CR10 CR10 Measurement and Control - Page 198

Grounding

Page 198 highlights

SECTION 14. INSTALLATION AND MAINTENANCE 14.7 GROUNDING 14.7.1 PROTECTION FROM LIGHTNING Primary lightning strikes are those where lightning hits the datalogger or sensors directly. Secondary strikes occur when the lightning strikes somewhere near the system and induces a voltage in the wires. The purpose of an earth ground is to minimize damage to the system by providing a low resistance path around the system to a point of low potential. Campbell Scientific recommends that all dataloggers in use in the field be earth grounded. All components of the system (datalogger, sensors, external power supplies, mounts, housings, etc.) should be referenced to one common earth ground. Every terminal on the Wiring Panel, with the exception of ground (G) and analog ground (AG)terminals are spark gapped. The spark gaps willfire at 150 V and the current will be diverted to ground. As shown in Figure 14.7-1, the power ground and analog ground are independent lines until joined inside the CR10. The fuse shown in Figure 14.7-1 (located on the underside of the Wiring Panel) is a 30 AWG wire, equivalent to a conventional 5 Amp fuse. It will blow if a sufficient transient comes in on the G or AG lines. at which time the current is SENSOR SHIELDS,+ REFERENCE FOR PULSE COUNTERS OR BINARY INPUTS RETURN FOR POWER TO EX1ERNAL DEVICES. directed away from the CR10 through the diodes. The fuse may be replaced by soldering another 30 AWG wire to the soldering pads provided. A modem/phone line connected to the Wiring Panel provides another path for transients to enter and damage the CR10. Campbell Scientific's DC112 phone modem has spark gaps on the phone lines. A 12 AWG wire should be run from the modem ground terminal to the earth ground. Additional protection is provided by the ground (Pin 2) of the 9 pin Serial l/O which is tied to power ground on the Wiring Panel. The transient protection designed into Campbell Scientific's equipment is meaningless if a good system earth ground is not provided. lt is the users responsibility to provide this earth ground. In laboratory applications, locating a stable earth ground is not always obvious. ln older buildings, new cover plates on old AC sockets may indicate that a safety ground exists when in fact the socket is not grounded. lf a safety ground does exist, it is good practice to verify that it carries no current. lf the integrity of the AC power ground cannot be verified, it is better to ground the system to a massive metalobject such as a steelwater pipe. WIRING PANEL +12V SINGLE ENDED REFERENCES AND EXCITATION RETURNS so ewc VREJ EQUIVALENT TO 5 AMP FUSE EXCITATION PORT EXCITATION VOLTAGE EXCITATION CONTROL EARTH GROUND FfGURE 14.7-1. Wiring PanelGrounding Diagram and Excitation Control 14-6

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159
  • 160
  • 161
  • 162
  • 163
  • 164
  • 165
  • 166
  • 167
  • 168
  • 169
  • 170
  • 171
  • 172
  • 173
  • 174
  • 175
  • 176
  • 177
  • 178
  • 179
  • 180
  • 181
  • 182
  • 183
  • 184
  • 185
  • 186
  • 187
  • 188
  • 189
  • 190
  • 191
  • 192
  • 193
  • 194
  • 195
  • 196
  • 197
  • 198
  • 199
  • 200
  • 201
  • 202
  • 203
  • 204
  • 205
  • 206
  • 207
  • 208
  • 209
  • 210
  • 211
  • 212
  • 213
  • 214
  • 215
  • 216
  • 217
  • 218
  • 219
  • 220
  • 221
  • 222
  • 223
  • 224
  • 225
  • 226
  • 227
  • 228
  • 229
  • 230
  • 231
  • 232
  • 233
  • 234
  • 235
  • 236
  • 237
  • 238

SECTION
14.
INSTALLATION AND
MAINTENANCE
14.7
GROUNDING
14.7.1
PROTECTION
FROM LIGHTNING
Primary lightning strikes are those where
lightning
hits
the datalogger or sensors directly.
Secondary strikes occur when
the
lightning
strikes
somewhere
near
the
system and
induces
a
voltage
in
the
wires.
The
purpose
of
an earth ground
is
to
minimize damage to
the
system by providing
a
low resistance path
around the system
to
a point
of
low potential.
Campbell
Scientific
recommends that
all
dataloggers in use
in
the
field
be earth
grounded.
All components
of
the
system
(datalogger, sensors, external power supplies,
mounts, housings, etc.) should
be
referenced
to
one common
earth
ground.
Every terminal on the
Wiring
Panel, with the
exception of
ground
(G) and analog ground
(AG)terminals
are spark
gapped.
The spark
gaps
willfire
at
150
V and the current will be
diverted
to
ground.
As shown in Figure
14.7-1,
the power ground and analog ground are
independent lines
until
joined
inside
the
CR10.
The fuse shown
in
Figure 14.7-1 (located on
the
underside
of
the
Wiring
Panel) is a 30 AWG
wire, equivalent
to
a
conventional
5
Amp fuse.
It
will blow
if a
sufficient transient comes
in
on
the G or
AG
lines. at which time the
current
is
directed away from
the
CR10 through the
diodes.
The
fuse
may
be
replaced by soldering
another 30 AWG wire to
the
soldering pads
provided.
A
modem/phone line connected to the Wiring
Panel provides another path for transients to
enter and damage
the
CR10.
Campbell
Scientific's DC112 phone modem has
spark
gaps
on
the
phone
lines.
A
12
AWG wire
should
be
run
from
the
modem ground terminal
to the earth
ground.
Additional protection
is
provided
by
the
ground (Pin 2)
of
the 9 pin
Serial
l/O
which
is
tied
to
power ground on the
Wiring
Panel.
The transient protection designed into Campbell
Scientific's equipment is meaningless if a good
system earth ground is not
provided.
lt
is
the
users responsibility
to
provide this earth ground.
In
laboratory applications, locating
a
stable
earth ground is
not
always
obvious.
ln
older
buildings, new cover plates on old AC sockets
may indicate that a safety ground exists
when
in
fact the socket is not
grounded.
lf
a
safety
ground does
exist,
it
is good practice to verify
that
it
carries
no
current.
lf
the
integrity
of
the
AC
power ground cannot
be
verified,
it is better
to ground the system to
a
massive
metalobject
such as a
steelwater
pipe.
EARTH
GROUND
FfGURE
14.7-1. Wiring
PanelGrounding
Diagram
and Excitation Control
SENSOR
SHIELDS,+
REFERENCE
FOR
PULSE
COUNTERS
OR
BINARY
INPUTS
RETURN
FOR
POWER
TO
EX1ERNAL
DEVICES.
WIRING PANEL
+12V
so
ewc
VREJ
EQUIVALENT
TO
5
AMP
FUSE
EXCITATION
VOLTAGE
EXCITATION
CONTROL
SINGLE ENDED
REFERENCES
AND
EXCITATION
RETURNS
EXCITATION
PORT
14-6