Campbell Scientific CR10 CR10 Measurement and Control - Page 183

Bridge Resistance, Measurements - conductivity program

Page 183 highlights

Radiation shielding must be provided when a junction box is installed in the field. Care must also be taken that a thermal gradient is not induced by conduction through the incoming wires. The CR10 can be used to measure the temperature gradients within the junction box. 13.5 BRIDGE RESISTANCE MEASUREMENTS There are 6 bridge measurement instructions included in the standard CR10 software. Figure 13.5-1 shows the circuits that would typically be measured with these instructions. In the diagrams, the resistors labeled R, would normally be the sensors and those labeled R1 would normally be fixed resistors. Circuits other than those diagrammed could be measured, provided the excitation and type of measurements were appropriate. With the exception of Instructions 4 and 8, which apply an excitation voltage then wait a specified time before making a measurement, all of the bridge measurements make one set of measurements with the excitation as programmed and another set of measurements with the excitation polarity reversed. The error in the two measurements due to thermal emfs can then be accounted for in the processing of the measurement instruction. The excitation is switched on 450 ps before the integration portion of the measurement starts and is grounded as soon as the integration is completed. Figure 13.5-2 shows the excitation SECTION 13. CRlO MEASUREMENTS and measurement sequence for Instruction 6, a 4 wire full bridge. Excitation is applied separately for each phase of a bridge measurement. For example, in Instruction 6, as shown in Figure 13.5-2, excitation is switched on for the 4 integration periods and switched off between integrations. lnstruction 8 measurement sequence consists of applying a single excitation voltage, delaying a specified time, and making a differential voltage measurement. lf a delay of 0 is specified, the inputs for the differential measurement are not switched for a second integration as is normally the case (Section 13.2). The result stored is the voltage measured. Instruction 8 does not have as good resolution or common mode rejection as the ratiometric bridge measurement instructions. lt does provide a very rapid means of making bridge measurements as well as supplying excitation to ci rcu itry req u i ring-diff erential measurements. This instruction does not reverse excitation. A 1 before the excitation channel number (1X) causes the channel to be incremented with each repetition. The output of lnstruction 8 is simply the voltage measurement. When 8 is used to measure a full bridge (same connections as Instruction 6 in Figure 13.5-1), the result is V1 which equals V, (R./(R3+R4) - R2l(R1+R2)). (ln other words, to make the output the same as Instruction 6, use a factor of 10004/* in the multiplier.) 13-17

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159
  • 160
  • 161
  • 162
  • 163
  • 164
  • 165
  • 166
  • 167
  • 168
  • 169
  • 170
  • 171
  • 172
  • 173
  • 174
  • 175
  • 176
  • 177
  • 178
  • 179
  • 180
  • 181
  • 182
  • 183
  • 184
  • 185
  • 186
  • 187
  • 188
  • 189
  • 190
  • 191
  • 192
  • 193
  • 194
  • 195
  • 196
  • 197
  • 198
  • 199
  • 200
  • 201
  • 202
  • 203
  • 204
  • 205
  • 206
  • 207
  • 208
  • 209
  • 210
  • 211
  • 212
  • 213
  • 214
  • 215
  • 216
  • 217
  • 218
  • 219
  • 220
  • 221
  • 222
  • 223
  • 224
  • 225
  • 226
  • 227
  • 228
  • 229
  • 230
  • 231
  • 232
  • 233
  • 234
  • 235
  • 236
  • 237
  • 238

Radiation shielding must be provided when
a
junction
box is installed
in
the
field.
Care
must
also
be
taken that
a
thermal
gradient is not
induced by conduction through
the
incoming
wires.
The
CR10
can be
used
to
measure
the
temperature
gradients within the
junction
box.
13.5
BRIDGE RESISTANCE
MEASUREMENTS
There are
6
bridge
measurement
instructions
included
in
the
standard
CR10
software.
Figure
13.5-1 shows the circuits that would
typically
be
measured with
these
instructions.
In
the
diagrams,
the
resistors labeled
R,
would
normally
be
the
sensors and
those
labeled
R1
would normally
be
fixed
resistors.
Circuits
other
than
those
diagrammed
could be
measured,
provided
the
excitation and
type
of
measurements
were
appropriate.
With
the exception of
Instructions 4 and
8,
which apply an excitation voltage then wait a
specified
time before
making
a
measurement,
all
of
the bridge
measurements make one set
of
measurements with
the
excitation as
programmed and
another
set
of
measurements
with the
excitation
polarity
reversed.
The
error
in
the
two
measurements due to
thermal
emfs
can
then
be accounted
for
in
the
processing of
the
measurement
instruction.
The
excitation
is
switched on
450
ps before
the
integration
portion
of
the
measurement starts
and
is
grounded as soon as
the
integration
is
completed.
Figure
13.5-2 shows the excitation
SECTION
13.
CRlO MEASUREMENTS
and
measurement sequence
for
Instruction 6,
a
4 wire
full
bridge.
Excitation is applied
separately for each phase of
a
bridge
measurement.
For
example, in
Instruction 6, as
shown
in
Figure 13.5-2,
excitation
is switched
on for
the
4 integration periods
and
switched off
between integrations.
lnstruction 8
measurement sequence consists
of applying
a
single excitation voltage, delaying
a specified
time, and
making a differential
voltage measurement.
lf
a
delay of 0
is
specified,
the
inputs for the differential
measurement
are
not switched
for
a second
integration
as
is normally the
case
(Section
13.2).
The
result stored
is
the voltage
measured.
Instruction
8
does
not have as good
resolution or
common
mode rejection as
the
ratiometric
bridge
measurement
instructions.
lt
does
provide
a
very
rapid means
of
making
bridge measurements as
well
as supplying
excitation to
ci
rcu
itry
req
u
i
ring-diff
erential
measurements. This
instruction
does
not
reverse
excitation. A
1
before the excitation
channel
number (1X) causes the channel
to
be
incremented with
each
repetition.
The
output
of
lnstruction
8
is simply the voltage
measurement. When 8
is used
to
measure
a
full bridge
(same
connections as
Instruction
6
in
Figure
13.5-1),
the
result
is V1
which equals
V,
(R./(R3+R4)
-
R2l(R1+R2)).
(ln
other words,
to
make
the
output
the
same
as
Instruction
6,
use
a
factor of 10004/*
in
the
multiplier.)
13-17