Campbell Scientific CR10 CR10 Measurement and Control - Page 188

Calibration

Page 188 highlights

SECTION 13. CRlO MEASUREMENTS ln Figure 13.6-2, V" is the excitation voltage, R1 is a fixed resistor, R, is the sensor resistance, and R6 is the resistance between the excited electrode and CR10 earth ground. With R6 in the network, the measured signal is: Rs + V1=V" [13.6-1] (Rr+R1) RrRy'R6 RsRy'Rc is the source of error due to the ground loop. When R6 is large the equation reduces to the ideal. The geometry of the electrodes has a great effect on the magnitude of this error. The Delmhorst gypsum block used in the 227 probe has two concentric cylindrical electrodes. The center electrode is used for excitation; because it is encircled by the ground electrode, the path for a ground loop through the soil is greatly reduced. Moisture blocks which consist of two parallel plate electrodes are particularly susceptible to ground loop problems. Similar considerations apply to the geometry of the electrodes in water conductivity sensors. The ground electrode of the conductivity or soil moisture probe and the CR10 earth ground form a galvanic cell, with the water/soil solution acting as the electrolyte. lf current was allowed lo flow, the resulting oxidation or reduction would soon damage the electrode, just as if DC excitation was used to make the measurement. Campbell Scientific probes are built with series capacitors in the leads to block this DC current. In addition to preventing sensor deterioration, the capacitors block any DC component from affecting the measu rement. 13.7 CALIBRATION PROCESS The CR10 makes voltage measurements by integrating the input signal for a fixed time and then holding the integrated value for the analog to digital (A/D) conversion. The A"/D conversion is made by a 13 bit approximation using a digital to analog converter (DAC). The result from the approximation is DAC counts, which are multiplied by coefficients to obtain millivolts (mV). There are 10 calibration coefficients, one for each of the 5 gain ranges for the fast and slow integration times. The CR10 has an internalcalibration function that leeds positive and negative voltages through the amplifiers and integrator and calculates new calibration coefficients. By 13-22 adjusting the calibration coefficients the accuracy of the voltage measurements is maintained over the -25 to +50'C operating range of the CR10. Calibration is executed under four conditions: 1. When the CR10 is powered up. 2. Automatically when Instruction 24 is not contained in a program table. 3. When the watchdog resets the processor. : 4. When the calibration instruction. Instruction 24, is executed. AUTOMATIC CALIBRATION SEQUENCE The primary advantage of automatic calibration is that the CR10 is constantly calibrated without user programming. The CR10 defaults to automatic calibration when lnstruction 24 is not contained in a program table. Every 8 seconds one part of a22part calibration sequence is performed. Program execution is interrupted (5.4 - 21.4 ms), when necessary, for each part of the calibration. Every 2.9 minutes (8 seconds * 22) len calibration coefficients are calculated. The calculated coefficients are multiplied by 1/5, and then added to 415 times the existing coetficients. Averaging is done as a safeguard against coefficients calculated from a noisy measurement. The above weighting of the newly calculated coetficients results in a 15 minute time constant (see Instruction 58) in the response of the calibration to step changes affecting the calibration coefficients (primarily temperature). For most environmental applications a 15 minute time constant is acceptable. The automatic calibration may result in the calibration coefficients not being optimum for applications that subject the CR10 to extreme temperature gradients. Automatic calibration extends the processing time 5.4 to 21.4 ms when it is executed (every 8 seconds). lf the processing time exceeds the execution interval the CR10 finishes processing the table and awaits the next occurrence of the execution interval before initiating the table. At the fastest execution interval of 1/64 (0.0156) second the program lable WILL be overrun by

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159
  • 160
  • 161
  • 162
  • 163
  • 164
  • 165
  • 166
  • 167
  • 168
  • 169
  • 170
  • 171
  • 172
  • 173
  • 174
  • 175
  • 176
  • 177
  • 178
  • 179
  • 180
  • 181
  • 182
  • 183
  • 184
  • 185
  • 186
  • 187
  • 188
  • 189
  • 190
  • 191
  • 192
  • 193
  • 194
  • 195
  • 196
  • 197
  • 198
  • 199
  • 200
  • 201
  • 202
  • 203
  • 204
  • 205
  • 206
  • 207
  • 208
  • 209
  • 210
  • 211
  • 212
  • 213
  • 214
  • 215
  • 216
  • 217
  • 218
  • 219
  • 220
  • 221
  • 222
  • 223
  • 224
  • 225
  • 226
  • 227
  • 228
  • 229
  • 230
  • 231
  • 232
  • 233
  • 234
  • 235
  • 236
  • 237
  • 238

SECTION
13.
CRlO
MEASUREMENTS
ln
Figure 13.6-2,
V"
is
the excitation
voltage,
R1
is a
fixed
resistor,
R,
is
the sensor
resistance,
and
R6
is
the
resistance between the excited
electrode and CR10 earth
ground.
With R6
in
the
network,
the
measured
signal
is:
Rs
V1=V"
[13.6-1]
(Rr+R1)
+
RrRy'R6
RsRy'Rc
is
the source of error due to the ground
loop.
When R6
is large the equation reduces
to
the
ideal.
The geometry
of
the
electrodes
has
a
great effect on
the
magnitude
of
this
error.
The
Delmhorst gypsum block used
in
the 227 probe
has
two concentric cylindrical
electrodes.
The
center
electrode
is used for excitation; because
it
is
encircled by
the
ground electrode,
the
path
for a ground loop through the soil
is
greatly
reduced.
Moisture blocks which consist of two
parallel plate electrodes are particularly
susceptible to
ground
loop
problems.
Similar
considerations apply to the geometry
of
the
electrodes
in
water conductivity sensors.
The ground electrode
of
the conductivity or soil
moisture probe and
the
CR10 earth ground
form
a
galvanic cell, with the water/soil solution
acting as the
electrolyte.
lf
current was allowed
lo
flow,
the
resulting oxidation
or
reduction
would soon damage the electrode,
just
as
if
DC
excitation
was
used
to
make
the
measurement.
Campbell Scientific probes are built
with
series
capacitors
in
the
leads
to
block
this
DC
current.
In
addition to preventing sensor deterioration,
the
capacitors
block
any
DC
component from
affecting
the
measu
rement.
13.7
CALIBRATION
PROCESS
The
CR10 makes
voltage
measurements by
integrating
the
input signal for
a
fixed time and
then
holding
the
integrated value for the analog
to digital (A/D)
conversion.
The
A"/D
conversion
is made by
a
13 bit approximation using a digital
to analog
converter
(DAC).
The
result from
the
approximation is DAC counts, which are
multiplied by coefficients to
obtain
millivolts
(mV).
There
are
10
calibration coefficients, one
for each
of
the
5
gain
ranges for the fast and
slow integration times.
The
CR10 has
an internalcalibration
function
that
leeds
positive
and
negative voltages
through the amplifiers and integrator and
calculates
new calibration
coefficients.
By
13-22
adjusting the calibration coefficients
the
accuracy
of
the
voltage
measurements
is
maintained over
the
-25 to
+50'C
operating
range
of
the
CR10.
Calibration
is
executed
under four conditions:
1.
When
the
CR10 is powered
up.
2.
Automatically
when
Instruction
24
is
not
contained
in
a
program table.
3.
When the
watchdog
resets
the
processor.
:
4.
When the calibration instruction. Instruction
24, is executed.
AUTOMATIC CALIBRATION
SEQUENCE
The
primary advantage of automatic calibration
is
that
the
CR10
is
constantly calibrated without
user
programming. The
CR10 defaults
to
automatic calibration
when
lnstruction
24
is not
contained in a program table.
Every 8 seconds one part
of
a22part
calibration sequence is
performed.
Program
execution
is interrupted (5.4
-
21.4
ms), when
necessary,
for
each part of the calibration.
Every 2.9 minutes
(8
seconds
*
22)
len
calibration coefficients are
calculated.
The
calculated coefficients
are
multiplied
by
1/5,
and
then added to
415
times the existing
coetficients. Averaging
is
done as a safeguard
against coefficients calculated
from a
noisy
measurement.
The above weighting
of
the
newly calculated
coetficients
results
in
a
15 minute time constant
(see Instruction
58)
in
the
response of
the
calibration to step changes affecting
the
calibration
coefficients
(primarily temperature).
For most environmental applications
a
15
minute time constant is
acceptable.
The
automatic
calibration
may result
in
the
calibration
coefficients
not being optimum
for
applications that subject
the
CR10 to extreme
temperature gradients.
Automatic calibration extends
the
processing
time 5.4
to
21.4 ms when it is executed (every
8
seconds).
lf
the
processing time exceeds
the
execution interval the CR10
finishes
processing
the table and awaits
the
next occurrence
of
the
execution interval before initiating the
table.
At
the fastest execution interval
of
1/64 (0.0156)
second
the
program
lable WILL
be
overrun
by